4. Information Readout Command - I**
Communication for RA2000 series (1WMPD4003507)
4-16
ICH (Inquire CHannel) Reading extra event (E1) setting
ICH (Inquire CHannel) Reading invalid amp setting
IUS (Inquire User Scale) Reading user-scale
Function
Outputs extra event (E1) setting.
ICH E1, P2 (Delimiter)
Input Format
P1: E1 fixed
P2: Signal number [1-16]
A1, A2, A3, A4, A5, A6, A7, A8 (Delimiter)
Output Format
A1: Amp type
A2: Input
A3: <Reserved>
A4: Signal ON/OFF
A5: EV wavelength position
A6: Vibration
A7: Width of base line
-1 fixed
(0=OFF, 1=ON)
(0=OFF, 1=ON) The order of all 16 signals
is sig1, 2, 3, to 16 from left.
[0.0 to 215.0 [mm]
2.0 to 25.0 [mm]
0.5 to 2.0 [mm]
Description
When an event unit is not installed, the output becomes the same as an
invalid amp so that “0,0,0,0”, is output.
Function
Outputs the value meaning that the selected channel is an invalid amp.
ICH P1 (Delimiter)
Input Format
P1: Selecting channel [1-16]
Output Format Outputs “0,0,0,0”.
Description
Function
Outputs user-scale
Input Format
IUS P1 (Delimiter)
P1: Selecting channel
(For RA2300MK
Ⅱ
, 1 through 16, for RA2800A, 1 through 32)
Output Format A1: ON,OFF for physical conversion(0=OFF, 1=ON)
A2: maximum input value(Can be omitted)
A3:minimum input value (Can be omitted)
A4:maximum output value (Can be omitted)
A5:minimum output value (Can be omitted)
A6:upper limit of recording full scale. (Can be omitted)
A7:lower limit of recording full scale. (Can be omitted)
A8:Unit setting(Can be omitted)
0= Standard, 2=N, 3=Pa, 4=mm, 5=µ
ε
, 6= m/s
2
,
7=
゚
C
, 8=
Ω
, 9= kg, 10= kgf, 11= kgf/cm
2
, 12= g
A9: User-specified unit (character string of a maximum of 9 characters)
(Can be omitted)
Description
When the selected channel is the amp other than an analog type of amp, a
parameter error occurs.
When an error occurs, "?,?,?,?,?,?,?,?,?" is returned.
Summary of Contents for Omniace RA2000 series
Page 1: ...Communication Command RA2000 series INSTRUCTION MANUAL 1WMPD4003507 ...
Page 2: ......
Page 3: ...Communication Command RA2000 series INSTRUCTION MANUAL ...
Page 4: ......
Page 14: ......
Page 15: ...1 Selection of Communication Interface ...
Page 27: ...2 Overview of Communication Control ...
Page 34: ......
Page 35: ...3 Setting Command s ...
Page 63: ...4 Information Readout Command I ...
Page 90: ......
Page 91: ...5 Execution Command E ...
Page 99: ...6 File Data Operation Command F ...
Page 102: ......
Page 103: ...7 Text Operation Command T ...
Page 109: ...8 Reference ...
Page 113: ......