11
8.
COMPARATOR
The scale has three-, five- and seven-level comparators.
Each comparator mode compares the weight value against the preset limit values and
outputs the results using LEDs (yellow / green / red).
Note: When the unit is "lb-oz" or "tl", this function can not be used.
Five-level comparator mode:
Uses four comparator values to
compare the weight value and
outputs results in five levels of
LOLO, LO, OK, HI and HIHI.
Three-level comparator mode:
Uses two comparator values (upper
and lower limit values) to compare the
weight value and outputs results in
three levels of LO, OK and HI.
Seven-level comparator mode
(portion weighing mode):
Uses six comparator values to
compare the weight value and
outputs results in seven levels
of outside the lowest limit, level 1 (LOLO), level 2 (LO), Level 3 (OK), level 4 (HI),
level 5 (HIHI) and outside the highest limit.
To use the comparator modes, the function settings
and
must be
specified and the comparator values must be set.
Using the function setting
, select a comparator mode.
0
: Five-level comparator mode (Result LED blinks)
1
: Five-level comparator mode (Result LED lights)
2
: Three-level comparator mode (Result LED blinks)
3
: Three-level comparator mode (Result LED lights)
4
: Seven-level comparator mode (Result LED blinks)
5
: Seven-level comparator mode (Result LED lights)
Using the function setting
, select comparison conditions.
0
: No comparison (comparator disabled).
1
: To compare all data.
2
: To compare all stable data.
3
: To compare all data which are
+5d or ≤ -5d.
4
: To compare stable data which are
+5d or ≤ -5d.
5
: To compare all data which are
+5d.
6
: To compare stable data which are
+5d.
d = minimum display in kg (Refer to "
14. SPECIFICATIONS
".)
In the counting mode, "d" is equal to the minimum weight display of kg mode.