12
7.2. Setting of the weighing capacity / minimum division / decimal point position
Set the weighing capacity, minimum division, and decimal point position. Be sure to
perform these setting first when connecting to a load cell (base unit). The indicator does
not require setting each time, if nothing is to be changed.
1
.
Refer to “7.1. Items of the calibration mode”, to enter the
calibration mode. The indicator displays “
Cal5et
”.
2. Press the TARE key to enter the setting mode of
the weighing capacity, minimum division, and decimal
point position.
3. The indicator displays the currently set minimum
division and calibration unit. (Display example: “
d-1
kg”) The blinking digit indicates the minimum
division. Press the PRINT key to select 1, 2, or 5.
The unit indicates the calibration unit. Press the
SAMPLE key to select either “kg” or “lb”.
Press the TARE key to store the currently set
minimum division and calibration unit, and proceed to
the next step.
4. After displaying “
Cap
”, the indicator displays the
currently set weighing capacity and decimal point
position. (Display example: “
0060.00
kg”) Set the
weighing capacity and decimal point position using
the following keys:
MODE
Shift the blinking digit to the right.
PRINT Increases the number (+1) at the
blinking digit.
SAMPLE Shift the decimal point position.
Press the TARE key to store the currently
displayed weighing capacity and decimal point
position, and proceed to the next step.
Available setting range of the resolution is 60,000 or
less. The resolution is the value that divides the
weighing capacity by the minimum division.
(Decimal point position is ignored.)
5. After displaying “
end
”, the indicator displays “
Cal5et
”,
and the setting
is
complete.
After setting the weighing capacity, minimum division, and decimal point position, be
sure to perform “7.4. Calibration using a weight”.
Press
Press
Press
Press