5
5. INSTALLATION and PRECAUTIONS
5.1. Precautions for Installing the Indicator
DANGER
Ground the indicator so that the user will not be subjected an electric shock.
Do not handle the main power cord with wet hands.
The AC plug is not water-resistant. Install it in an area where it does not get
wet.
Even when the optional sealed lead acid (SLA) battery (HC-02i) is used, be
sure to prevent the AC plug from getting wet.
Do not install the indicator where there is flammable or corrosive gas present.
Do not pull, fold or arrange cables forcibly.
Use shielded cables for all connections.
Ground the platform to be connected to the indicator to avoid a risk of electric
shock.
Read the section “PRECAUTIONS” in the instruction manual of the SW to
understand the conditions and precautions for installation, operation and cleaning
the indicator.
5.2. Connecting to the load cell
1. Disconnect the main power cord of the scale from the power source.
2. On the bottom of the display, remove the screw protection caps on the load cell
panel. Loosen the four screws and open the load cell panel.
3. Loosen the cable clamp locking nut on the load cell panel and insert the load cell
cable through the cable clamp.
Accommodates load cell cable diameters of
φ
3.5 to 7.0 mm.
Locking
nut
Load cell
cable
Load cell
panel
Cable
clamp