41
Five-level comparator mode
…
Set the LO limit value.
Three-level comparator mode
…
Not used
Seven-level comparator mode
…
Set the level 3 lower limit value.
(No reply for the function setting “
aCk 0
“.)
Send the six-digit value excluding the polarity and decimal point.
Command
L 1
,
+ 0 0 0 2 0 0 C
R
L
F
Reply
L 1
,
+ 0 0 0 2 0 0 C
R
L
F
Five-level comparator mode
…
Set the LOLO limit value.
Three-level comparator mode
…
Set the LO limit value.
Seven-level comparator mode
…
Set the level 2 lower limit value.
(No reply for the function setting “
aCk 0
“.)
Send the six-digit value excluding the polarity and decimal point.
Command
L 2
,
+ 0 0 0 1 0 0 C
R
L
F
Reply
L 2
,
+ 0 0 0 1 0 0 C
R
L
F
Five-level comparator mode
…
Not used
Three-level comparator mode
…
Not used
Seven-level comparator mode
…
Set the level 1 lower limit value.
(No reply for the function setting “
aCk 0
“.)
Send the six-digit value excluding the polarity and decimal point.
Command
L 3
,
+ 0 0 0 0 0 0 C
R
L
F
Reply
L 3
,
+ 0 0 0 0 0 0 C
R
L
F
Replies to the commands other than examples above when the function setting “
aCk 1
“ is
selected
.
The scale is not in a state where a command could be executed. Then, the scale will reply “I”.
Command
Z C
R
L
F
Reply
I C
R
L
F
The scale is not in a condition that zero operation is possible.
Command does not exist for the scale. Then, the scale will reply “?”.
Command
B C
R
L
F
Reply
? C
R
L
F
The scale received an undefined command.
When the function setting “
aCk 0
“ is selected, undefined commands are ignored
and no reply is sent.