20
10.1. The formula to compare
Comparison is performed using the formula listed below and the results are output.
Five-level comparator mode
Results
Comparison formula
LED display
LOLO
Displayed value < LOLO limit value
(Or over in the negative value)
( Red LED on)
LO
Displayed value < LO limit value
( Yellow LED on)
OK
LO limit value
≤
Displayed value
≤
HI limit value
( Green LED on)
)
HI
HI limit value < Displayed value
( Yellow LED on)
HIHI
HIHI limit value < Displayed value
(Or over in the positive value)
( Red LED on)
Three-level comparator mode
Results
Comparison formula
LED display
LO
Displayed value < LO limit value
(Or over in the negative value)
( Red LED on)
OK
LO limit value
≤
Displayed value
≤
HI limit value
( Green LED on)
)
HI
HI limit value < Displayed value
(Or over in the positive value)
( Red LED on)
Seven-level comparator mode (portion weighing mode)
Results
Comparison formula
LED display
None
Displayed value < Level 1 lower limit value
(Or over in the negative value)
( No LEDs on)
LOLO
(Level 1)
Displayed value < Level 2 lower limit value
( Red LED on)
LO
(Level 2)
Displayed value < Level 3 lower limit value
( Yellow LED on)
OK
(Level 3)
Level 3 lower limit value
≤
Displayed value
≤
Level 3 upper limit value
( Green LED on)
)
HI
(Level 4)
Level 3 upper limit value < Displayed value
( Yellow LED on)
HIHI
(Level 5)
Level 4 upper limit value < Displayed value
( Red LED on)
None
Level 5 upper limit value < Displayed value
(Or over in the positive value)
( No LEDs on)
The comparator values are common to the weighing and counting mode.
Ignore the decimal point when setting the comparator values.
For example, when the weighing capacity is 6 kg and the setting value is “001000”:
Display mode
Limit value
Capacity / Minimum display
Normal resolution kg
1.000 kg
6.000 kg / 0.002 kg
High resolution kg
1.000 kg
6.000 kg / 0.001 kg
Maximum resolution kg
0.1000 kg
6.0000 kg / 0.0005 kg
Normal resolution oz
100.0 oz
210.0 oz / 0.1 oz
High resolution oz
10.00 oz
210.00 oz / 0.05 oz
Maximum resolution oz
10.00 oz
210.00 oz / 0.02 oz
Counting mode
1000 pcs
The comparator values are maintained even if the power is turned OFF.
Judgment order of comparison is from the top row to the bottom in the comparator
mode tables
The entered comparator values are not judged. Even if the upper limit value is less
than the lower limit value, no error will be output.