English  8

 

 When a measurement is made with an 

erratic pulse or in a very noisy 

environment, 

 

will display. 

 

7.  Turning the power off 
 Press 

the 

 START 

  button to turn the power off.   

The device will be in standby mode. 

8. Subsequent 

measurements 

 

If a subsequent measurement is required, turn off the power and turn it on 
again. When "0 " is displayed, the device is ready for another measurement. 

Note:  Allow at least three minutes between measurements on the same 

person. 

9.  Automatic power-off function 

 

If the device is left on after a measurement, it will turn off automatically 

after about one minute. It may be turned off at any time by pressing the 

 

START 

 button. 

 

Notes For Proper Measurement 

 Sit down in a comfortable position. Place the arm to be used for the 

measurement on a table or other support so that the center of the cuff will be 
at the same height as your heart. 

 Relax for about five or ten minutes before taking a measurement. If you are 

excited or depressed by emotional stress, the measurement will reflect this 
stress as a higher (or lower) than normal blood pressure reading and the 
pulse reading will usually be faster than normal. 

  Your blood pressure varies constantly, depending on what you are doing and 

what you have eaten. What you drink can have a very strong and rapid effect 
on your blood pressure. 

 This device bases its measurements on the heartbeat. If you have a very 

weak or irregular heartbeat, the device may have difficulty determining your 
blood pressure. 

 Should the device detect a condition that is abnormal, it will stop the 

measurement and display an error symbol. Refer to page 4 for the 
descriptions of the symbols. 

 This blood pressure device is intended for use by adults only. Consult with 

your physician before using this device on a child. A child should not use this 
device unattended. 

 
 
 
 
 
 
 

 

Summary of Contents for UA-705

Page 1: ...Digital Blood Pressure Monitor Model UA 705 Instruction Manual Manuel d instructions Manual de Instrucciones Manuale di Istruzioni 1WMPD4000988B...

Page 2: ...Measurement Operation 6 After Measurement 6 Measurements 7 Measurement With Storing Data 7 Notes For Proper Measurement 8 Operation Mode 9 WHO Classification Indicator 9 Pressure Bar Indicator 9 What...

Page 3: ...use The device is for use indoors Precautions Precision components are used in the construction of this device Extremes in temperature humidity direct sunlight shock or dust should be avoided Clean t...

Page 4: ...English 3 Parts Identification...

Page 5: ...surement until exhaust is complete Press the exhaust valve button to exhaust air from the cuff Flashes when pressure is insufficient for the measurement Pump air into the cuff with the rubber bulb Pre...

Page 6: ...attery if the device is not to be used for a long time The battery may leak and cause a malfunction The stored data are cleared when the batteries are removed Connecting The Air Hose Insert the air co...

Page 7: ...right after physical exercise or a bath Rest for twenty or thirty minutes before taking the measurement Try to measure your blood pressure at the same time every day Measurement Operation During measu...

Page 8: ...hen the mark disappears Note If you want to stop inflation at any time stop squeezing the rubber bulb and press the exhaust valve button to expel the air from the cuff 4 When inflation is complete def...

Page 9: ...ter of the cuff will be at the same height as your heart Relax for about five or ten minutes before taking a measurement If you are excited or depressed by emotional stress the measurement will reflec...

Page 10: ...ain WHO Classification Indicator Each of the six segments of the bar indicator corresponds to the WHO blood pressure classification described on the next page Pressure Bar Indicator The indicator moni...

Page 11: ...h arterial blood pressure if left unattended can cause many health problems including stroke and heart attack Hypertension can be controlled by altering one s lifestyle avoiding stress and with medica...

Page 12: ...Battery polarities are not in the correct position Reinstall the battery with its negative and positive end matching those indicated in the battery compartment The cuff is not fastened properly Fasten...

Page 13: ...functioning and accuracy Please contact either your authorized dealer or A D for maintenance Technical Data Type UA 705 Measurement method Oscillometric measurement Measurement range Pressure 20 280...

Page 14: ......

Reviews: