9. CLEANING 

1.  Turn the power off and unplug from the wall outlet. 

2.  Remove the air tube from the air inlet connectors. 

3. 

TO CLEAN: 

Remove the mouthpiece, nosepiece, mask, and disassemble the 

nebulizer cup (cup, cap, and baffle), and wash these items in hot water with 
dish-washing detergent. Rinse these items thoroughly to remove the detergent 
and air dry. 

4. 

TO DISINFECT: 

Mix one part of white vinegar with 3 parts hot water in a 

clean container. Submerge mouthpiece, nosepiece, mask, and parts of 
nebulizer cup for half an hour in the solution. Remove from solution and let 
air dry. 

5.  There is no need to clean the tube. If necessary wipe the surface regularly. 

Note: 

  Replace the nebulizer accessories including adult mask, child mask, 

mouthpiece, nosepiece, nebulizer cup, Air tube after using 6 months. 

  Replace the filter every 30 days or if the filter turns grey. 

 

CAUTION: The device must be changed if it becomes clogged.  

CAUTION: The device and accessories must not be boiled. 

 
 

10. MAINTENANCE 

1.  The air filter should be replaced when it turns gray. Additional filters can be 

procured from your dealer. 

2.  Only authorized personnel shall make repairs to this device. 

 

CAUTION: The device may be damaged if used with a dirty filter or if 

the air filter is replaced by some other material such as 

cotton. Do NOT operate without the dedicated air filter.

Summary of Contents for UN-019

Page 1: ...pressor Nebulizer Model UN 019 Instruction Manual Original Manuel d instructions Traduction Manual de Instrucciones Traducci n Manuale di Istruzioni Traduzione Bedienungsanleitung bersetzung 1WMPD4004...

Page 2: ......

Page 3: ...RODUCTION Page 5 6 SPECIFICATIONS Page 5 7 PARTS IDENTIFICATION Page 6 8 OPERATING INSTRUCTIONS Page 7 9 CLEANING Page 8 10 MAINTENANCE Page 8 11 TROUBLESHOOTING Page 9 12 TECHNICAL DATA Page 10 13 EM...

Page 4: ...r respiratory disorders This device should be used only as directed Do not use this device for anything other than as prescribed by your doctor Before use read this instruction manual thoroughly 2 PRE...

Page 5: ...his device is intended for aerosol therapy only Any other use is not recommended 5 Do not twist the air tube during operation 6 Do not pour more than 6ml of solution into the device DANGER Risk of ele...

Page 6: ...ing in wet or damp locations 10 Disconnect the power cord from the electrical outlet before filling the nebulizer cup 11 When using this device near TVs microwave ovens pulse telephones X rays or othe...

Page 7: ...psi 241 kPa to 345 kPa Neb Operating Pressure 9 16 psi 62 kPa to 110 kPa Liter Flow Range 5 8 L min Average Nebulization Rate Approx 0 33 ml min 6 ml medicine can be nebulized within 18 minutes Maxim...

Page 8: ...lug UN 019A EC1B BF plug UN 019A EC2B Replacement parts list Appearance Part name Part code Qty Material Adult mask UNB 014A 01EC 1 PVC Child mask UNB 014A 02EC 1 PVC Air tube UNB 014A 03EC 1 PVC Air...

Page 9: ...on the nebulizer cup is the maximum level Medication cannot be filled beyond the maximum level 6 Insert the mouthpiece into the top of the nebulizer cup If nosepiece is used insert it into the mouthpi...

Page 10: ...let airdry 5 There is no need to clean the tube If necessary wipe the surface regularly Note Replace the nebulizer accessories including adult mask child mask mouthpiece nosepiece nebulizer cup Air t...

Page 11: ...le is free of blockages The nebulizer cup is tilted at a sharp angle Make sure that the nebulizer kit is not tilted at an angle of more than 45 degrees The air tube is not connected correctly Make sur...

Page 12: ...s for particle size MMAD Mass Median Aerodynamic Diameter 13 EMD TECHNICAL DATA Medical Electrical Equipment needs special precautions regarding EMD and needs to be installed and put into service acco...

Page 13: ...3 See table 4 Rated power frequency magnetic fields IEC 61000 4 8 30 A m 50 Hz or 60 Hz Table 3 IMMUNITY TEST LEVELS Input a c power Port Phenomenon IMMUNITY TEST LEVELS Electrical fast transients bu...

Page 14: ...5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz 2 0 3 28...

Page 15: ...rm to the environment or human health from uncontrolled waste disposal please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources...

Page 16: ...English 14 MEMO...

Page 17: ...44 1235 550485 A D ENGINEERING INC 4622 Runway Boulevard Ann Arbor MI 48108 USA Telephone 1 888 726 9966 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8...

Reviews: