background image

INSTRUCTIONS TO USER

Dear Users, thank you very much for purchasing the UP-200 Pulse Oximeter.

In case of modifications and software upgrades, the information contained in 

this document is subject to change without notice.

The Manual describes, in accordance with the Pulse Oximeter’s features and 

requirements, product features, functions, specifications, correct methods for:  

transportation, use, repair, maintenance and storage, as well as the safety 

procedures to protect both the user and equipment. Refer to the respective 

chapters for details.

Please read the Manual very carefully before using this equipment. These 

instructions describe the operating procedures to be followed strictly, failure to 

follow these instructions can cause measuring abnormality, equipment damage 

and personal injury. The manufacturer is NOT responsible for the safety, 

reliability and performance issues and any monitoring abnormality, personal 

injury and equipment damage due to user’s negligence of the operation 

instructions. The manufacturer’s warranty service does not cover such faults. 

Due to product updates, the specific products you received may not be 

totally in accordance with the description of this User Manual. We regret any 

confusion. 

This product can be used repeatedly. Its useful life is 3 years. 

If you have any questions regarding to the use of this product, please call us at 

one of the numbers listed at the end of this manual.

WARNING:

• An uncomfortable or painful feeling may appear if using the oximeter 

continuously, especially for the microcirculation barrier users. It is 

recommended that the sensor should not be applied to the same 

finger for over 2 hours.

• For individual users, ensure proper placement of the oximeter. The 

oximeter cannot be clipped on the edema and tender tissue.

• The light (the infrared is invisible) emitted from the oximeter is harmful 

to the eyes, do not stare at the light.

• User cannot use enamel or other fingernail polish.

• User’s fingernail cannot be too long.

• Please review the relative content about the clinical restrictions and 

caution.

• This oximeter is not intended for treatment.

The User Manual is published by our company. All rights reserved.

1 SAFETY

1.1 Instructions for Safe Operations

•  Check the main unit and all accessories periodically to make sure that 

there is no visible damage that may affect user’s safety and monitoring 

performance. It is recommended that the oximeter should be inspected at 

least once a week. When there is obvious damage, stop using the oximeter.

•  Necessary maintenance must be performed by qualified service engineers 

ONLY. Users are not permitted to maintain it by themselves.

•  The oximeter cannot be used together with devices not specified in User’s 

Manual. Only use the type of accessories included in the package.

•  This product is calibrated before leaving factory.

1.2 Warnings

•  Explosive hazard—DO NOT use the oximeter in environment with 

inflammable gas such as some ignitable anesthetic agents.

•  Do not use this oximeter if you are allergic to rubber.

•  The disposal of the oximeter and its accessories and packaging (including 

batteries, plastic bags, foam inserts and paper boxes) should follow the 

local laws and regulations.

•  Please check the packing before use to make sure the oximeter and 

accessories are in accordance with the packing list, or else the oximeter 

may not work properly. 

•  Please don’t measure this oximeter with a functional tester for the 

oximeter’s related information. 

•  Warning against servicing and maintenance while the equipment is in use. 

•  No modification of this equipment is allowed.

•  The user is an intended operator.

•  The probe of the oximeter is the applied part.

1.3 Attentions

•  Keep the oximeter away from dust, vibration, corrosive substances, 

explosive materials, high temperature and moisture.

•  If the oximeter gets wet, discontinue use.

•  Do not use immediately after moving it from a cold environment to warm or 

humid environment. 

•  Do not operate buttons on front panel with sharp materials.

•  Do not use high temperature or high pressure steam disinfection of the 

oximeter. Refer to chapter (8) for instructions of cleaning and disinfection.

•  Do not immerge the oximeter in liquid. When it needs cleaning, please wipe 

its surface with medical alcohol by soft material. Do not spray any liquid on 

the oximeter directly.

•  When cleaning the oximeter with water, the temperature should be lower 

than 60 °C.

•  As to the fingers which are too thin or too cold, it would probably affect the 

normal measure of the users’ SpO

2

 and pulse rate, use a larger finger such 

as thumb or middle finger and place finger deeply into the probe.

•  Do not use the oximeter on infant or neonatal users.

•  The product is suitable for adults (weight should be between 40 kg to 110 

kg).

•  The oximeter may not work for all users. If you are unable to achieve stable 

readings, discontinue use.

•  The update period of data is less than 5 seconds, which is changeable 

according to different individual pulse rate. 

•  The waveform is normalized. Please read the measured value when the 

waveform on the screen is steady. This measured value is optimal value, 

and the waveform at the moment is the standard one.

•  If some abnormal conditions appear on the screen during test process, pull 

out the finger and reinsert to restore normal use.

•  The lanyard is made from non-allergy material, if you are sensitive to the 

lanyard, stop using it. In addition, do not wear it around the neck to avoid 

harm to the users.

•  The oximeter does not have low-voltage alarm function, it only shows the 

low-voltage. Please change the battery when the battery energy is used up.

•  The instrument does not have alarm function. Do not use the device in 

situations where alarms are required. 

•  Batteries must be removed if the oximeter is going to be stored for more 

than one month, or else batteries may leak.

•  A flexible circuit connects the two parts of the oximeter. Do not twist or pull 

on the connection.

1.4 Indication for Use

The Pulse Oximeter is a non-invasive device intended for the spot-check of 

saturation of arterial hemoglobin(SpO

2

) and the pulse rate of adult in home use 

environments. This oximeter is not intended for continuous monitoring. The 

device can be used multiple times.Pulse oximeter intended for wellness use.

2 OVERVIEW

The pulse oxygen saturation is the percentage of HbO

2

 in the total Hb in the 

blood - the O

2

 concentration in the blood. It is an important bio-parameter for 

respiration. This Pulse Oximeter was developed for the purpose of measuring 

the SpO

2

 more easily and accurately. At the same time, the oximeter can 

measure the pulse rate.

The Pulse Oximeter features a compact design, low power consumption, 

and convenient operation. It is only necessary for user to put one finger into 

a fingertip photoelectric sensor for measurement, and a display screen will 

directly show measured value of Hemoglobin Saturation.

2.1 Features

•  Easy to use

•  Easy to view with a display that changes direction automatically.

•  Small, lightweight design

•  Convenient carrying case.

•  Low power consumption – allowing for 20 hours of continuous operation on 

a new set of batteries.

•  Standby mode when no signal is received within 5 seconds.

2.2 Major Applications and Scope of Application

The Pulse Oximeter can be used in measuring pulse oxygen saturation and 

pulse rate through finger. The product is suitable for family use (It can be used 

before or after doing sports, and it is not recommended to use the oximeter 

during sports activity).

The problem of overrating would emerge when the user is 

suffering from toxicosis which caused by carbon monoxide, the 

oximeter is not recommended to be used under this 

circumstance.

2.3 Environment Requirements

Storage Environment

a.  Temperature: -40°C ~ +60°C

b.  Relative humidity: ≤95%

c.  Atmospheric pressure: 500hPa ~ 1060hPa

Operating Environment

a.  Temperature: 10°C ~ 40°C

b.  Relative humidity: ≤75%

c.  Atmospheric pressure: 700hPa ~ 1060hPa 

3 ACCESSORIES

•  1 Lanyard

•  2 Batteries

•  1 User Manual

•  1 Carrying Case

4 SET UP

4.1 View of the Front Panel

pulse bar graph

pulse waveform

Low-battery indication SpO

2

pulse rate

button

Figure 1: Front View

4.2 Battery

Step 1.Insert the two AAA size batteries in the proper direction (Refer to Figure 

2).

Step 2. Put the cover back on.

Please take care when you insert the batteries for the improper 

insertion may damage the device. 

Figure 2: Batteries Installation

4.3 Attaching the lanyard

Step 1. Put the thin loop through the hole on the oximeter.

Step 2. Put the lanyard strap through the thin loop and tighten.

Figure 3: Attaching the Lanyard

Figure 4: Put Finger in Position

5 USING THE OXIMETER

5.1  Insert the two batteries properly to the direction, and then replace the 

cover.

5.2  Open the clip as shown in Figure 4. 

5.3  Place your finger into the rubber cushions with the nail towards the top of 

the oximeter, and then clip your finger.

5.4  Press the switch button once on front panel.

5.5  Do not shake your finger and remain still.

5.6  See your results from screen display.

5.7  The button has two functions. When the oximeter is in standby mode, 

pressing the button can exit it; When the oximeter is in operation status, 

pressing the button for longer can change brightness of the screen.

5.8  The oximeter changes display direction according to the direction you 

hold it.

Fingernails and the luminescent tube (emission tube) should be 

on the same side.

6 PRINCIPLE AND CAUTION

6.1 Principle of Measurement

Principle of the Oximeter is as follows: An experience formula of data process 

is established taking use of Lambert Beer Law according to Spectrum 

Absorption Characteristics of Reductive Hemoglobin (Hb) and Oxyhemoglobin 

(HbO

2

) in glow & near-infrared zones. Operation principle of the instrument is: 

Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance 

with Capacity Pulse Scanning & Recording Technology, so that two beams 

of different wavelength of lights can be focused onto human nail tip through 

perspective clamp finger-type sensor. Then measured signal can be obtained 

by a photosensitive element, information acquired through which will be shown 

on screen through treatment in electronic circuits and microprocessor.

Glow and Infrared-ray 

Emission Tube

Glow and Infrared-ray 

Receipt Tube

Figure 5: Operating Principle

6.2 Caution

1.  The finger should be placed properly (see the attached illustration of this 

manual, Figure 4), or else it may cause inaccurate measurement.

2.  The SpO

2

 sensor and photoelectric receiving tube should be arranged in a 

way with the subject’s arteriole in a position there between.

3.  The SpO

2

 sensor should not be used at a location or limb tied with arterial 

canal or blood pressure cuff or receiving intravenous injection.

4.  Make sure the optical path is free from any optical obstacles like rubberized 

fabric.

5.  Excessive ambient light may affect the measuring result. It includes 

fluorescent lamp, dual ruby light, infrared heater, direct sunlight etc.

6.  Strenuous action of the subject or extreme electrosurgical interference may 

also affect the accuracy.

7.  User cannot use fingernail polish.

6.3 Clinical Restrictions

1.  As the measurement is taken on the basis of arteriole pulse, substantial 

pulsating blood flow of subject is required. For a subject with weak pulse 

due to shock, low ambient/body temperature, major bleeding, or use of 

vascular contracting drug, the SpO

2

 waveform (PLETH) will decrease. In 

this case, the measurement will be more sensitive to interference.

2.  For those with a substantial amount of staining dilution drug (such as 

methylene blue, indigo green and acid indigo blue), carbon monoxide 

hemoglobin (COHb), methionine (Me+Hb) or thiosalicylic hemoglobin, and 

some with icterus problem, the SpO

2

 determination by this monitor may be 

inaccurate.

3.  Drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also 

cause inaccurate SpO

2

 measurements. 

4.  As the SpO

2

 value serves as a reference value for anemic anoxia and 

toxic anoxia, some users with serious anemia may report good SpO

2

 

measurement.

7 TECHNICAL SPECIFICATIONS

1. 

Display Format:

 LCD Display 

SpO

2

 Measurement Range: 

0 ~ 100% 

Pulse Rate Measuring Range:

 30 bpm ~ 250 bpm 

Pulse Wave Display:

 bar-graph display and the waveform display.

2. 

Power Requirements: 

2 × 1.5 V AAA alkaline battery (or rechargeable 

battery), adaptable range: 2.6 V ~ 3.6 V.

3. 

Power Consumption: 

Less than 30 mA.

4. 

Resolution:

 1% for SpO

2

 and 1 bpm for Pulse Rate.

5. 

Measurement Accuracy:  

SpO2: ±2% for 70% ~ 100%; not applicable if < 70% 

Pulse Rate: ±2 bpm or ±2% (whichever is larger) 

Clinical Trial: SpO

2

 regression plot & Bland–Altman plot,Refer to Figures 

6 and 7.

6. 

Measurement Performance in Weak Filling Condition: 

SpO

2

 and pulse 

rate can be shown correctly when pulse-filling ratio is 0.4%. SpO

2

 error is 

±4%, pulse rate error is ±2 bpm or ±2% (select larger).

7. 

Resistance to surrounding light: 

The deviation between the value 

measured in the condition of man-made light, indoor natural light and 

darkroom is less than ±1%.

8. 

It is equipped with a switch function: 

The product will enter standby 

mode when no signal is in the product within 5 seconds.

9. 

Optical Sensor 

Red light (wavelength is 660 nm, 6.65 mW)  

Infrared (wavelength is 905 nm, 6.75 mW)

Figure 6: SpO

2

 regression plot

Figure 7: Bland–Altman plot

8 REPAIRING AND MAINTENANCE

•  Please change the batteries when low-voltage is displayed on the screen.

•  Clean the surface of the oximeter before using. Wipe the oximeter with 

medical alcohol first, and then let it dry in air. 

•  Using the medical alcohol to disinfect after use prevents cross infection for 

next use.

•  Please take out the batteries if the oximeter is not used for a long time.

•  The oximeter can be transported by ordinary conveyance or according to 

transport contract. The oximeter can not be transported mixed with toxic, 

harmful, corrosive material.

•  The best storage environment of the oximeter is - 40ºC to 60ºC ambient 

temperature and not higher than 95% relative humidity. 

•  There is no need to calibrate the oximeter.

High-pressure sterilization cannot be used on the oximeter. 

Do not immerse the oximeter in liquid.

It is recommended that the oximeter should be kept in a dry 

environment. Humidity may reduce the useful life of the oximeter, 

or even damage it.

Pulse Oximeter User Manual

UP-200 

Reviews: