Warnings related to use of the Darvall Cozy

TM

 Warm Air Heater

 

 

Grounding reliability can ONLY be achieved when the Darvall Cozy Heater’s power cord is 

connected to a properly grounded Mains power outlet. 

If the over-temperature indicator light illuminates and the alarm sounds, discontinue using of 

Darvall Cozy Heater to prevent thermal injury to the patient. Unplug the Darvall Cozy Heater 

and contact a qualified service technician. 

Do not use the Darvall Cozy Heater if it is not connected to blankets or Thermal Injury may 

result.  The Cozy Heater is designed to be connected to a purpose-made semi-permeable 

forced warm air blanket that distributes warm air evenly around the animal. 

Caution using Darvall Cozy Warm Air Heating System in animals with transdermal 

medication. Drug delivery may be increased which could result in animal injury or death. 

Do not allow the animal to lie on Darvall Cozy Heater Hose or allow the hose to directly contact 

the animal’s skin during warming, or thermal injury may occur.  

When  replacing  the  Darvall  Cozy  Heater Hose,  always  disconnect  the  temperature  sensor 

attached to the old hose. Be sure to attach the hose and temperature sensor from the new 

hose, prior to using the device. 

Do not use Darvall Cozy Heater in MRI rooms as the heater contains ferrous metal and emits 

RF interference. Darvall can supply all-plastic heater extension hoses to duct warm air from 

outside the MRI room or the magnetic field to warm air blankets covering animals in MRI  

The Darvall Cozy Heater meets medical electronic interference requirements (IEC60601-1-2) 

Clamping the Darvall Cozy Heater to an IV pole stand can result in instability. To prevent this, 

use the Darvall Cozy Heater no higher than 60 cm (24") high on an IV pole stand. 

Do not use Cozy Heater replacement components in that are not supplied by Darvall/AAS or 

their distributors or authorized service agents 

Electrical shock hazard: Do not disassemble the Darvall Cozy Heater unless you are a 

qualified service technician. There are electrically live parts in the Darvall Cozy Heater if it is 

connected to a power source, even when it not switched on OR is in Standby mode. 

End-users may replace the Darvall Cozy HEPA Air Filter, the heater hose with temperature sensor, 

the heater hose cover and the electrical fuse located near the mains supply inlet power cord. 

 

 

 

Summary of Contents for DARVALL Cozy

Page 1: ......

Page 2: ...Regional Customer Support...

Page 3: ...ing to use Darvall Cozy Warm Air Heater 8 Installation of the Darvall Cozy Wheels 9 Use the Darvall Cozy Heater with an appropriate warming blanket 10 Darvall Cozy Warm Air Heater Control Panel Indica...

Page 4: ...rmly to caged and anaesthetized animals International Patents granted and applied for The full range of Darvall Warm Air Blankets for use in cages anesthesia or surgery are shown on Page 6 The Darvall...

Page 5: ...always disconnect the temperature sensor attached to the old hose Be sure to attach the hose and temperature sensor from the new hose prior to using thedevice Do not use Darvall Cozy Heater in MRI ro...

Page 6: ...6...

Page 7: ...is mounted on 4 swivel wheels Figure 2 Darvall Cozy Heater Rear view The power switch is next to the Power Cord inlet and will disconnect all power to the Darvall Cozy if in the OFF position The fuse...

Page 8: ...alified service technician in accordance with the instructions in the Service Manual Warranty repairs must be performed by AAS Darvall or be approved in writing if undertaken by a third party qualifie...

Page 9: ...the floor onto the wheels Removal of Wheels from the Darvall Cozy Heater Tools Adjustable Spanner Time 5 min Removing the 40mm swivel wheels each with its push fit wheel mount adapter is essentially...

Page 10: ...se soiling Attach the Darvall Cozy heater hose to the Warming Blanket Insert the hose nozzle into the warming blanket s entry port so as to minimize kinks at this point Wrap the blanket tapes firmly r...

Page 11: ...he air Temperature set point is set to room temperature Air flow button When selected the air flow will be toggled between high and low speed LED on Low Air Flow Mode LED off High Air Flow Mode 3 Powe...

Page 12: ...ater willheat HEPA filtered air to MIDDLE temp setting 38 C 4 43 C a Press 43 C HIGH temperature button b The LED light above the button will illuminate c The Darvall Cozy Warm Air Heater willheat HEP...

Page 13: ...emperature alarms sound if the temperature limits are exceeded for example if the Darvall Cozy Heater is set to warm to 43 C and the Heater Hose outlet temperature sensor 1 is outside of 43 3 C or Coz...

Page 14: ...s 1 Fan stopped 2 Alarm prompts Repeat interval of 10 seconds E03 PTC ERR Low Pitch Trill 3 rings 1 PTC heating element stopped working 2 Alarm prompts Repeat interval of 10 seconds E04 OVER TEMPERATU...

Page 15: ...rvall Cozy Warm Air Heater and wait for 5 minutes 2 Restart the Darvall Cozy 3 If the display shows the same fault code contact a qualified service technician to replace the PTC heater E04 Over Temper...

Page 16: ...mal Infection Risk using warm air heating during surgery in an appropriate manner For any sterile procedure use of a new Darvall Warm Air Blanket is advisable In human surgery compared to no warming t...

Page 17: ...ull tabs of the HEPA filter and lift out It should pull out easily 8 Open new filter box and take out the HEPA air filter and Filter Changed sticky label 9 Install the new HEPA filter as show with Pul...

Page 18: ...removal and installation of a new hose the temperature sensor wire MUST BE disconnected old hose and re connected new hose If Temperature Sensor 2 is not connected the Darvall Cozy Heater WON T RUN 8...

Page 19: ...t down operating indicator lights turn OFF control panel becomes unresponsive Fault Fault information will display on the LCD alarm sounds Over current protection Dual input fused lines Electrical Cha...

Page 20: ...erating at one of the 4 operating temperatures to switch the Darvall Cozy into Standby mode press the ON OFF button Operating Mode When entering operating mode the following events occur Unit will att...

Page 21: ...airs but also has a loaner program so NO downtime is required Shipping Procedures Repairs on all products will be accepted with authorisation numbers However the following information is required lett...

Reviews: