33
9.5 Technical safety testing
WARNING - Technical safety tests may only be performed by the manufacturer, the AAT field
service or an authorised specialist dealer!
WARNING - It is recommended to perform a technical safety test every two years!
The test badge shows when the test is to be carried out at the latest! Tip: After reuse / change
of user!
9.6 Reuse (change in operator)
The term reuse is used when the AAT product of patient X is to be used for another patient Y. Please
contact the manufacturer, the AAT field staff or the authorised specialist shop.
The c-max U1 is suitable for reuse. A technical safety test and cleaning/disinfection work should be
performed prior to each reuse. (Refer to chapters 9.3 and 9.5)
i
NOTE - You may not mend or bypass a damaged fuse!
Only new fuses with
30 A
may be replaced!
i
NOTE - If the fuse blows frequently, contact the manufacturer or an authorised specialist dealer!
i
NOTE - Only fuses purchased from the specialist dealer or AAT may be used as replacement fuses!
Remove the fuse cover on
the battery pack.
» To do this, loosen the screw.
Remove covers and old fuses
and insert new fuses.
» Replacement fuses are
available from the manu-
facturer!
Position fuse cover and fix it
with the screw.
9.4 Replacing fuses
To ensure flawless device function, we recommend carrying out a safety check
every 2 years. We also recommend the technical safety test before each reuse.
Training is required in order to be able to perform the technical safety testing. For
more information, please visit: www.aat-online.de