V9000T Trinocular Microscope    

 

 

 
 

 

 

SAFETY NOTES 

 

1.  Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from 

dropping and being damaged. 

 

2.  Do not discard the molded Styrofoam container; the container should be retained should the 

microscope ever require reshipment. 

 

3.  Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.  

Ensure the microscope is located on a smooth, level and firm surface. 

 

4.  If any specimen solutions or other liquids splash onto the stage, objective or any other component,  

disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be 
damaged. 

 

5.  All electrical connectors (power cord) should be inserted into an electrical surge suppressor to 

prevent damage due to voltage fluctuations. 

 

6.  For safety when replacing the LED bulb 

or fuse, be sure the main switch is off (“O”), remove

 the 

power cord, and replace the LED bulb after the bulb and the lamp house has completely cooled. 

 

7.  Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The 

use of a different input voltage other than indicated will cause severe damage to the microscope. 

 
 

 

CARE AND MAINTENANCE

 

1.  Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly. 

 

2.  Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces 

should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap 
solution. Do not use organic solvents for cleansing. 

 

3.  The outer surface of the optics should be inspected and cleaned periodically using an air stream from 

an air bulb. If dirt remains on the optical surface, use a soft cloth or cotton swab dampened with a 
lens cleaning solution (available at camera stores). All optical lenses should be swabbed using a 
circular motion. A small amount of absorbent cotton wound on the end of a tapered stick such as 
cotton swabs or Q-tips, makes a useful tool for cleaning recessed optical surfaces. Avoid using an 
excessive amount of solvents as this may cause problems with optical coatings or cemented optics 
or the flowing solvent may pick up grease making cleaning more difficult.  Oil immersion objectives 
should be cleaned immediately after use by removing the oil with lens tissue or a clean, soft cloth. 

 

4.  Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not 

in use. 

 

5.  ACCU-SCOPE

®

 microscopes are precision instruments which require periodic preventative 

maintenance to maintain proper performance and to compensate for normal wear. An annual 
schedule of preventative maintenance by qualified personnel is highly recommended. Your 
authorized ACCU-SCOPE

 ® 

distributor can arrange for this service. 

 

 

Summary of Contents for V9000T

Page 1: ...Operator s Manual V9000T Trinocular Microscope ...

Page 2: ...rinocular Microscope Original User s Manual For Veterinary Use Only Customer and Technical Support For USA 1 800 822 2947 For Europe 49 6155 780 210 techsupport abaxis de April 2016 PN 790 7013 Rev D Text 2016 Abaxis Inc Union City CA 94587 ...

Page 3: ......

Page 4: ... ASSEMBLY DIAGRAM PROCEDURE 6 8 ADJUSMENT OPERATION ILLUMINATION 9 MIRROR 9 SPECIMEN SLIDE 9 FOCUSING 10 INTERPUPILLARY DISTANCE 10 FIELD DIAPHRAGM 11 APERTURE DIAPHRAGM 11 OIL OBJECTIVE 12 FILTER 13 FUSE REPLACEMENT 13 LED BULB REPLACEMENT 14 TECHNICAL SPECIFICATIONS 16 TROUBLESHOOTING 17 19 MAINTENANCE 20 SERVICE 20 20 WARRANTY ...

Page 5: ...ves or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persistent dirt should be removed using a mild soap solution Do not use organic solvents for cleansing 3 The outer surface of the optics should be inspected and cleaned periodically using an air stream from an air bulb If dirt remains on ...

Page 6: ...ality prior to shipment UNPACKING AND COMPONENTS Do not discard the container Your microscope arrived packed in a molded Styrofoam container the Styrofoam container should be retained for reshipment of your microscope if needed Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew will form Carefully remove the microscope from the Styrofoam con...

Page 7: ... Trinocular Microscope 5 COMPONENTS DIAGRAM Lock Screw Y Axis Adjustment Knob Coarse Focusing Knob Fine Focusing Knob X axis Adjustment Knob Power Switch Iris Diaphragm Adjustment Control Objective Nosepiece ...

Page 8: ...assembled by our factory technicians at our New York facility prior to shipment Should you need to disassemble assemble your microscope in the future please follow the instructions outlined below When assembling the microscope make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces Eyepiece Objective Condenser Power Cord ...

Page 9: ...n the condenser lock screw Insert the condenser into the hole of the stand according to the directional arrow until the condenser is even with the stand and then rotate the condenser to make sure the iris diaphragm control lever faces forward Tighten the lock screw of the condenser then raise the condenser with the up down knob to the highest position Objectives Rotate the coarse focusing knob to ...

Page 10: ... power cord securely into the power socket see Fig 3 securely into an Insert the other connector electrical outlet The microscope uses a universal power supply so it may be used in any voltage range between 90 240v when used with the appropriate line cord An optional cord hanger can be installed on your microscope for simple and easy storage see Photos 1 2 Fig 3 Photo 1 Optional Cord Hanger Photo ...

Page 11: ...er switch to the 0 off position Loosen the Allen screw with an Allen wrench to screw off the built in illuminator housing see Fig 5 Screw on the mirror according to the directional arrow and tighten the Allen screw with the Allen wrench Rotate the mirror stand to fill the field with light Placing the Specimen Slide Push the slide finger of the specimen holder backwards Place the slide into the sli...

Page 12: ...r left eye only and the 10X objective focus your specimen by adjusting the coarse adjustment knob When the image is in view refine the image to its sharpest focus by turning the fine adjustment knob Rotate the diopter collar to obtain the sharpest focus To obtain the same sharp image using your right eye do not touch the coarse or fine adjustments Instead rotate the right diopter collar until the ...

Page 13: ...lluminator Condenser Optional Adjusting the Aperture Diaphragm The aperture size is increased or decreased by rotating the condenser aperture diaphragm lever When the aperture is closed the brightness and resolution are decreased but the contrast and range of focus are increased If the aperture diaphragm is opened the brightness and resolution are increased however the contrast and range of focus ...

Page 14: ...il immersion objective is in the light path With your eye at the level of the stage use the coarse focus knob to raise the stage with the specimen cover glass When you see a flash of light at this location the objective lens has made contact with the immersion oil and the microscope can now be focused using the fine focus knob Each time you finish using the oil immersion objective wipe off all tra...

Page 15: ... side downward onto the groove of the illuminator condenser Place the cover back on the illuminator condenser and re install the set screw Fig 13 Filters are available in clear blue frosted blue green yellow and white Replacing the Fuses Turn the main switch to 0 OFF before replacing a fuse Unplug the power cord from the microscope Screw off the fuse or from the fuse base with a Phillip s head typ...

Page 16: ...the main switch to 0 OFF before replacing the LED bulb Unplug the power cord from the microscope Gently place the microscope on its arm as shown in Figure 15 Using a Philips head screwdriver remove the 4 screws to release the base plate Figure 16 Disconnect the make break connector Figure 17 ...

Page 17: ...in position and screw it to its mount Figure 19 Figure 18 above Reinstall the base plate and reconnect the make break connector with the connector from the new bulb Figure 20 Secure the base plate with the 4 Philips screws Figure 21 Figure 21 NOTE With the new bulb installed the condenser may need to be re centered To adjust the condenser centration please follow the Assembly instructions for the ...

Page 18: ...imit tension adjustment Travel range 25mm Fine focusing precision 0 002mm Stage Built in low position coaxial control mechanical stage area 140x132mm moving range 76x50mm Condenser Built in Koehler illuminator systems Pre centered Iris diaphragm Koehler illuminator condenser and mirror optional Illuminator Universal power supply 90 240V voltage 3W LED light with variable intensity control Operatio...

Page 19: ...y No cover glass attached to the slide Cover glass is too thick or thin Slide maybe upside down Immersion oil is on a dry objective especially the 40xR No immersion oil used with100xR objective Air bubbles in immersion oil Condenser aperture is closed or open too much Condenser is positioned too low Attach a 0 17mm cover glass Use a cover glass of the appropriate thickness 0 17mm Turn slide over s...

Page 20: ...Insufficient brightness Lamp intensity is too low Aperture diaphragm closed too far Condenser position too low Adjust the light intensity by rotating the intensity control dial and or iris diaphragm Open to the proper setting Position the condenser slightly lower than the upper limit MECHANICAL PROBLEMS Image will not focus with high power objectives Slide upside down Cover glass is to thick Turn ...

Page 21: ...electrical power Lamp bulb burnt out Fuse blown out Check power cord connection Replace bulb Replace fuse Slippage of focus when using the coarse focusing knob Tension adjustment is set too low Increase the tension on the focusing knobs Fine focus is ineffective Tension adjustment is set too high Loosen the tension on the focusing knobs ...

Page 22: ... workmanship for a period of five years from the date of invoice to the original end user purchaser The LED lamp is warranted for a period of two years from the date of invoice to the original end user purchaser This warranty does not cover damage caused in transit misuse neglect abuse or damage resulting from improper servicing or modification by other then ACCU SCOPE approved service personnel T...

Page 23: ...016 1300 7015 Rev A Abaxis Inc Animal Health Worldwide Headquarters 3240 Whipple Road Union City CA 94587 United States of America Tel 510 675 6500 www abaxis com Abaxis Europe GmbH Bunsenstrasse 9 11 64347 Griesheim Germany Tel 49 6155 780 210 Tel 800 822 2947 ...