I N
| O P EN C H A N N EL A N D O P EN TA N K M O U N T I N G K I T | I N/A N A I N S T/0 49 - EN R E V. A
4 Kit installation
Referring to Figure 3:
1
Pass M3 screw
A
(supplied) through the end of sensor chain
mounting adaptor
B
, feed it through the last link on
opposite end of chain
C
and screw M3 screw
A
into tapped
hole in opposite side of sensor chain mounting adaptor
B
.
Lock into position using supplied lock nut
D
.
2
Pass EZLink
™
sensor cable connector
E
and sensor cable
F
through the hole in sensor chain mounting adaptor
B
and
feed entire cable
F
and chain
C
length through.
3
Screw sensor body
G
onto sensor chain mounting
adaptor
B
using PTFE tape to seal.
NOTICE
Do not overtighten.
4
When the required chain/cable length has been determined,
secure the chain onto second threaded stud using locknut
(supplied).
NOTICE
Ideally, the measuring end of the sensor should be a
minimum of 300 mm (12 in) below the free surface of the
liquid being monitored.
5
Gently lower the sensor into the sample until the sensor is
fully immersed to the required depth.
6
Connect the EZLink sensor cable connector to the AWT420
transmitter.
WARNING
Ensure the cable/chain run is routed securely and safely and
does not present a trip hazard.
7
Secure the chain to the installation. Make sure that it does
not come in contact with any wall or solid object.