36
Electromagnetic Flowmeter Converter
6.3.2 Testing the Metering System
1) If the flow direction does not agree with the flow direction indicators in the display, then the flow direction indicators are be reversed.
Select the “Flow Direction” parameter in the “Operating Mode” submenu” and change accordingly.
!
WARNING:
Contact hazards exist when the housing cover is
removed and the power is turned on !
These tasks should only be performed by trained
personnel.
Does the supply power agree with the values listed on the
converter Instrument Tag?
No
Solder correct jumpers for the required voltage,
see Fig.20.
Yes
Are the converter and the flowmeter primary installed in the
proper locations ? (Flowmeter primary, Protection Class,
converter, temperature, vibration, cable length, cable type,
display indication in the correct flow direction).
No
Check allowable installation conditions, temperature
(insulate high temperature design flowmeter primary and
pipeline), Protection Class, vibration.
Yes
Does the wiring agree with the specifications in the
Interconnection Diagrams ?
No
Check the connections per the Interconnection Diagrams
beginning on Page 7.
Yes
Is the flowmeter primary correctly grounded?
No
Check the grounds per the Interconnection Diagram, also
see the Flowmeter Primary Instruction Bulletin.
Yes
Is the supply voltage value at terminals L & N within the
nominal voltage limits of +10/-10 % ?.
No
Provide supply voltage within the limits.
Yes
Is the flowmeter primary filled with fluid?
No
Fill pipeline.
Yes
Is switch S901 on the connection board correctly set for the
flowmeter primary type and size?
No
See Figs.18-20. Switch S901 is open for use with flowmeter
primaries 10DS2110/3110. Switch S901 closed for use with
flowmeter primaries 10DI1425
≥
20” [DN 500], 10DS3111A/
..B/..C
≥
20” [DN 500] and 10D1422 1/8”- 40” [DN3 to 1000].
Yes
Turn on the power. Does the display indicate 0 % ? Is the
correct flow direction indication
F for forward,
R for
reverse displayed in the 1st line ?
No
Adjust system zero according to instructions in 3.2. Check if
the flow direction indicators agree with the flow direction
1)
.
Fuse defective ? Conductivity too low ? To check see
Flowmeter Primary Instruction Bulletin.
Yes
For continuous flow has the Standard operating mode been
selected ? For pulsating flows has the Piston Pump
operating mode been selected?
No
Select the parameter Standard or Piston Pump in the
operating mode submenu. Turn on software filter, set
damping >2.4 sec.
Yes
Have the meter size, engineering units, flow range,
density, pulse factor etc. been entered?
No
Enter meter size, units and flow range, select current output
e.g. 0-20/4-20 mA.
Yes
Does the present flowrate value agree with the displayed
value and the output signals ?
No
If the flow indication and current output are noisy a software
filter can be turned on or the damping setting increased.
Conduct Function Test 5.4 or check converter with
Flowmeter Primary Simulator 6.2 .
Yes
The metering system is operational. Store the meter
location parameters in the external EEPROM. The meter
location parameters can also be recorded on the page 52 in
the Overview Page for future reference.
→
←
Summary of Contents for 50SM1000
Page 4: ......