14
°
H
D. Org
8.5 mg/l C
or
Diss. Organics
mg/l
Alarm 1 Setpoint
1.5 mg/l
Alarm 2 Setpoint
8.6 mg/l
Manual Clean
No
Yes
Lamp Disabled
No
Sensor
Calibration
Yes
…5
PROGRAMMING
5.2
Operating Display Page
Note.
This is the default page. The programme returns to this point from any of the programming pages if no data has
been entered after four minutes.
Dissolved Organics Mode Operating Page
Display in
inferred units
(see Appendix)
Absorbance/metre; colour °H; Coag. Dose mgl
–1
; mgl
–1
T.O.C.
Equivalent value in mgl
–1
Dissolved Organics
.
This is for display only. See
Set Up Outputs Page
for programming details.
This is for display only. See
Set Up Outputs Page
for programming details.
Switching the Lamp On/Off
In the interests of safety it is essential that the lamp is switched off before
doing any maintenance on the sensor. When off,
Lamp Disabled
is
displayed in the lower window of the
Diss. Organics
page; the top
window will be blank.
Press the
switch to disable the lamp and
to switch the lamp back on.
Manual Cleaning
Press the
switch to change
No
to
Yes
and press the
switch to start
a manual clean.
Return to top of
Operating Display Page
.