background image

A DS 4 20

  |  D I SS O LV ED OX YG EN SEN S O R  |  O I/A DS 4 2 0 - EN R E V. B

15

6  Diagnostics

The table below shows sensor-specific icon types, diagnostic 
messages and possible causes/suggested remedial action.

Note .

•  The diagnostic icons in the following tables conform to 

NAMUR 107.

•  For transmitter-specific diagnostics messages, refer to  

OI/AWT420.

NAMUR 107 diagnostic icons 

Failure

Check  

function

Out of 

specification

Maintenance 

required

Process value 
invalid due to 

malfunction in 

sensor or 

transmitter

Process value 

temporarily invalid 

due to specified 

action

Process value 

unreliable due to 

sensor operating 

outside specified 

range/limits

Process value valid 

– intervention 

required to correct 

operational 

conditions

S(n): PT failure

Diagnostic message

S(n): PT failure

Cause

The measurement taken from the temperature sensor is invalid 
indicating that the temperature sensor has failed, or the 
associated connections are either open-circuit or short-circuit.

Recovery action

Visually inspect the sensor/temperature sensor for signs of 
damage. A damaged sensor must be replaced.

Cycle the power to the transmitter.

If the failure persists, contact the local service organization.

S(n): D.O. Conc Meas Error

Diagnostic message

S(n): D.O. Conc Meas Error

Cause

The DO sensor cannot provide a valid dissolved oxygen 
measurement due to a sensor error.

Recovery action

Cycle the power to the sensor.

If the failure persists, contact the local service organization.

S(n): %Sat Meas Error

Diagnostic message

S(n): %Sat Meas Error

Cause

The DO sensor cannot provide a valid % saturation 
measurement due to a sensor error.

Recovery action

Cycle the power to the sensor.

If the failure persists, contact the local service organization.

S(n): Calibrating

Diagnostic message

S(n): Calibrating

Cause

Sensor calibration in progress.

Recovery action

This diagnostic clears when the calibration is complete.

S(n): Internal Comms Error

Diagnostic message

S(n): Internal Comms Error

Cause

The DO sensor cannot provide a valid dissolved oxygen 
measurement due to an error in the communication between 
the sensor and the SmartCap.

Recovery action

Make sure that the SmartCap is installed and correctly 
seated.

S(n): Cap Removed

Diagnostic message

S(n): Cap Removed

Cause

The SmartCap has been removed from the sensor or is not 
being recognized. A valid measurement cannot be made.

Recovery action

Make sure that the SmartCap is installed correctly.

Summary of Contents for ADS420

Page 1: ...ional EZClean compressed air supply unit connected to an ABB adapter nozzle or T piece a schedule can be configured to provide automatic in situ cleaning of the sensor For more information Further publications and publications for the associated AWT420 and AWT440 transmitters are available for free download from www abb com measurement or by scanning this code Search for or click on Data Sheet ADS...

Page 2: ...libration 100 and 0 saturation 10 Abort calibration 12 Calibration timings 12 Stability period 12 Slow sensor response 12 Calibration troubleshooting 12 Slow sensor calibration or no response to dissolved oxygen changes 12 5 Maintenance 13 Cleaning the SmartCap 13 Cleaning the lens 13 Cleaning the sensor body 14 6 Diagnostics 15 S n PT failure 15 S n D O Conc Meas Error 15 S n Sat Meas Error 15 S ...

Page 3: ... below Direct current supply only This symbol identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment This symbol indicates the need for protective eye wear This symbol indicates the need for protective hand wear Recycle separately fr...

Page 4: ...tablish and maintain any appropriate measures such as but not limited to the installation of firewalls application of authentication measures encryption of data installation of anti virus programs etc to protect the product the network its system and the interface against any kind of security breaches unauthorized access interference intrusion leakage and or theft of data or information ABB and it...

Page 5: ...ner Note The SmartCap serial number is programmed on the memory chip inside the SmartCap D Sensor body including temperature sensor E 10 m 39 3 ft fixed cable ADS420 cable detached ADS420 sensor components cable detached are shown in Figure 2 Figure 2 ADS420 sensor components cable detached Table 2 Sensor components cable detached Part Description A EZLink digital sensor connector B Shroud C Smart...

Page 6: ...type None Low Medium High None Salinity Unit Select the required salinity units PSU Practical Salinity Units or ppt parts per thousand PSU Salinity Correction Required when monitoring water containing high quantities of dissolved salts enter the required value between 0 and 42 Practical Salinity Units PSU leave at the default value of 0 PSU if salinity correction is not required 0 PSU Pressure Uni...

Page 7: ...n After calibration and sensor replacement reset the currents outputs and alarms see Calibrate menu below Calibrate menu Calibrate Menu Exit Select Used to calibrate the sensor Access to the Calibrate menu is permitted via the Calibrate and Advanced levels only Note During calibration current outputs and alarms are set to Hold automatically if Hold Outputs is enabled see below Menu Comment Default...

Page 8: ...ement producing false low readings after calibration If condensation occurs remove the sensor thoroughly dry the sensing element probe and thermistor before performing the calibration procedure Ensure the calibrations are performed in a vented but enclosed chamber to prevent pressure building up in the container Minimize ambient airflow across the sensor head during the calibration 5 At the transm...

Page 9: ...t Span Cal Place in 100 Sat Abort Continue 12 Press the key below the Continue prompt The Calibrate page is displayed Calibrate PV 98 26 Tmp 25 22 C Setting Please Wait Abort 13 When span calibration is complete a results page is displayed Calibrate PV 98 98 Tmp 24 90 C Offset 0 045 Slope 1 054 Exit In the event of a calibration failure the message Calibration Failed is displayed with an indicatio...

Page 10: ...nsation to form on the surface of the sensing element producing false low readings after calibration If condensation occurs remove the probe thoroughly dry the sensing element probe and thermistor before performing the calibration procedure Ensure the calibrations are performed in a vented but enclosed chamber to prevent pressure building up in the container Minimize ambient airflow across the sen...

Page 11: ...n 100 Sat Abort Continue 12 Press the key below the Continue prompt The Calibrate page is displayed Calibrate PV 98 26 Tmp 25 22 C Setting Please Wait Abort 13 When span calibration is complete the Calibrate Start Zero Cal page is displayed Calibrate Start Zero Cal Place in 0 Sat Abort Continue 14 Remove the sponge wafer from the calibration pot 15 Fill calibration chamber with fresh sodium sulfit...

Page 12: ...perating Instructions OI AWT420 The Calibration Log displays the message Cal Aborted refer to OI AWT420 for Calibration Log entries Calibration timings Stability period The temperature and concentration readings are monitored for up to 10 minutes until a stable reading is achieved When stability is achieved the calibration slope and offset values are determined Slow sensor response If the output f...

Page 13: ... Figure 3 Cleaning the SmartCap Cleaning the lens 1 Isolate the transmitter and remove the sensor from the process Referring to Figure 4 2 Unscrew the shroud A from the sensor body B 3 Grip the SmartCap C between your thumb and forefinger and carefully slide it off the sensor D NOTICE Do not touch the end of the SmartCap C Do not allow moisture or atmospheric humidity inside SmartCap C Do not wet ...

Page 14: ...ate the transmitter and remove the sensor from the process Referring to Figure 5 2 Gently scrub the shroud A and the sensor body B with a soft bristled brush or a nylon dish scrubber If it is necessary use a grease remover to remove grease or other matter from the sensor body or soak in vinegar and deionized water to remove mineral deposits or extensive fouling Figure 5 Cleaning the sensor body ...

Page 15: ... 3 If the failure persists contact the local service organization S n D O Conc Meas Error Diagnostic message S n D O Conc Meas Error Cause The DO sensor cannot provide a valid dissolved oxygen measurement due to a sensor error Recovery action 1 Cycle the power to the sensor 2 If the failure persists contact the local service organization S n Sat Meas Error Diagnostic message S n Sat Meas Error Cau...

Page 16: ...sor Warm up Diagnostic message S n Sensor Warm up Cause The DO sensor is warming up and cannot provide a valid DO measurement until the internal electronics have stabilized Recovery action This diagnostic clears when the internal electronics have stablized If the diagnostic persists contact the local service organization S n Sensor Warning Diagnostic message S n Sensor Warning Cause PV measured fr...

Page 17: ...nostic message S n Replace Cap Cause The SmartCap in the DO sensor reaches its expiry date within 4 weeks After its expiry date the SmartCap may continue to operate but its accuracy degrades and cannot be guaranteed to meet specification Recovery action 1 Replace the SmartCap ...

Page 18: ...n AWT4009010 AWT4009050 AWT4009100 AWT4009150 AWT4009250 AWT4009500 1 m 3 3 ft 5 m 16 4 ft 10 m 32 8 ft 15 m 49 2 ft 25 m 82 0 ft 50 m 164 0 ft SmartCap spare parts Part number Description 3KXA494400L0019 SmartCap replacement kit Shroud spare parts Part number Description 3KXA494400L0020 Shroud replacement kit ...

Page 19: ...ADS420 DISSOLVED OX YGEN SENSOR OI ADS420 EN REV B 19 Acknowledgements EZLink is a trademark of ABB Limited ...

Page 20: ...h regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden with...

Reviews: