35
13
SPECIFICATION
4234 Power Supply Unit
Approvals
CENELEC approved
[EEx ia] IIC T
amb
–20
C to +55
C
BASEEFA Certificate No. BAS 01 ATEX 7041
II (1)G
Input Voltage
115V AC 50/60Hz
(4234501) or
230V AC 50/60Hz
(4234500)
Fuse Rating
250mA HRC ceramic
DC Output
350mA stabilized
0.14%
Load Conditions
1 Katharometer
13
:
max.
Interconnecting cable
2
:
max.
Ambient Temperature Range
–20 to 55
C (–4 to 131
F)
Supply Variations
15V (115V supply) or
30V (230V supply) 46 to 64Hz
Regulation
Within
0.5% for:
Load variation of
15%
Supply variation of
15%
Ambient temperature variation of
20
C (36
F)
4Hz frequency variation
Ripple
Less than 0.5% of set output peak/peak across
a 10
:
load
Stability
Within
0.7% of initial setting, over period of 1 month with load
resistance, supply voltage and ambient temperature at nominal
stated values
Overall Dimensions
160 x 170 x 110mm (6.3 x 6.7 x 4.3 in.)
Weight
2.12kg (4.8 lb) approx.
Environment
Sheltered interior
6540–203 and 6548–000
Katharometer Analyzer Panel
Approvals
CENELEC approved
EEx ia IIC T
amb
–20
C to +55
C
BASEEFA Certificate No. BAS 01 ATEX 1042
II (1)G
Model 6540–203
incorporating Model 6539–960 (H
2
) or
Model 6539–960 (CO
2
) Katharometer Unit
Model 6548–000
incorporating Model 6548–001
(H
2
and CO
2
) Katharometer Unit
Power Supply
350mA DC, from 4234500 or 4234501 power supply unit
Signal Output
0 to 10mV for each range
Accuracy
2% of scale span, each range
Dead Time
Typically 5s
Response Time
Typically 40s for 90% step change at Katharometer
Tubing and drying chamber introduce extra delays
Ambient Temperature
55
C (131
F) max.
Sample Connections
Compression couplings:
6mm OD tube (Model 6548-000)
8mm OD tube (Model 6540-203)
Sample Pressure
Minimum 125mm H
2
O
Maximum 0.35bar (Gauge)
Model 6540–203
Maximum 10bar (Gauge)
Model 6548–000
Normal Sample Flowrate
100 to 150ml/min.
Maximum Gas Flowrate
250ml/min
Minimum Gas Flowrate
50ml/min
Outline Dimensions
610 x 305 x 152mm (24 x 12 x 6 in.)
Weight
8.6kg (18.9lb)
Environment
Sheltered interior
Note.
Variations to the earlier certificate (SFA 3012:1972) allow the use of
these items in systems supplied to that standard.
SS/AK100 Issue 2