AquaMaster 3 FET200
Electromagnetic flowmeter Transmitter
2 Mechanical Installation
OI/FET200–EN Rev. I
13
7.
Slide SIM card
D
into holder
C
, contact side down and bevelled edge to the top-right.
8.
Close holder
C
until it clicks into place and refit cover
B
.
9.
Screw cover
B
firmly in place.
10. Refit the transmitter to its mounting point.
11. Position the transmitter on the integral mount and secure with 4 x 4 mm Hex fixing screws
A
.
Tighten the screws to 4Nm.
Note.
Step 10 is applicable only to remote- and close-coupled transmitters.
Note.
Step 11 is applicable only to integral transmitters.
Fig. 2.17 Installing a SIM Card
B
C
D
A