14 Sensor setup
2-electrode conductivity
Menu
Comment
Default
Tag
Enter an alphanumeric sensor tag (16 characters maximum) to identify the sensor on the
Operator Pages
.
TAG1
Measurement Type
Select measurement type:
• Conductivity/Concentration/Resistivity
Note
. If a change is made the I/O sources are reset.
Conductivity
Note .
The following menus are displayed only if measurement type =
Conductivity
Conductivity Unit
Select the conductivity units:
• mS/cm/μS/cm
μS/cm
Cell Constant
Enter the cell constant of the measuring cell used –
see the relevant conductivity cell manual.
1.00
Range High
Set the span value used in
Chart
and
Bargraph
views.
Cell constant dependent
– see table below
Range Low
Set the zero value used in
Chart
and
Bargraph
views.
0
Note .
The following menus are displayed only if measurement type =
Concentration
Cell Constant
Enter the cell constant of the measuring cell used –
see the relevant conductivity cell manual.
N/A
Concentration Unit
Select the concentration units:
• None(Blank)/ppm/mg/l/ppb/μg/l/%/Custom
N/A
Custom Units
Note
. Displayed only if concentration units =
Custom
Enter an alphanumeric string (6 characters maximum) for the custom (user defined) concentration units.
N/A
Conc . Curve Table
Set the user defined concentration curve using the 6-point linearizer table
(concentration against conductivity).
N/A
Range High
View the span value used in
Chart
and
Bargraph
views.
N/A
Range Low
View the zero value used in
Chart
and
Bargraph
views.
N/A
Filter Type
Select the signal filtering type:
• None/Low/Medium/High
None
Temp . Comp . Type
Set the type of temperature compensation:
• Manual/Automatic/None
Automatic
Manual Temperature
Note .
Displayed only if temperature compensation type =
Manual
Enter the temperature of the sample within the range –10.0 to 120.0 °C.
25.0 °C
TC Curve
Note .
Not displayed only if temperature compensation type =
None
Set the temperature compensation characteristic required:
• TC Coeff./Standard KCl/UPW (Low TC)/UPW (High TC)/Pure H20 (Neutral)/
Pure H2O (Acid)/Pure H2O (Base)/NaOH/HCl/NaCl/NH3/User Defined
TC Coeff
User Def . TC Curve
Note .
Displayed only if temperature compensation curve =
User Defined
Set the user defined temperature compensation curve using the six point linearizer table (% against °C).
N/A
TC Coefficient
Note .
Displayed only if temperature compensation curve =
User Defined
Enter the temperature coefficient (α x 100) of the solution (0.01 to 5.00 %/°C).
If unknown, the temperature coefficient (α) of the solution must be calculated – see page 72.
2.00 %/°C
Sensor Diagnostics
Polarisation
Detect excessive polarisation condition:
• Enabled/Disabled
Disabled
Out Of Solution
Detect
Out Of Solution
condition:
• Enabled/Disabled
Disabled
Reset To Defaults
Select to reset all
Sensor Setup
parameters to their default values.
Conductivity cell constant
Conductivity measuring range
0.01
0 to 200 µS/cm
0.05
0 to 1000 µS/cm
0 to 1 mS/cm
0.10
0 to 2,000 µS/cm
0 to 2 mS/cm
1.00
0 to 20,000 µS/cm
0 to 20 mS/cm