64
AW T4 20 |
U N I V ER S A L 4 - W I R E , D UA L- I N P U T T R A N SM I T T ER | O I/AW T4 2 0 - EN R E V. B
…15 Calibration procedures
… pH/Redox/ORP calibration
1-Point automatic calibration
Note
. Before starting the calibration ensure the pH buffers
are set to the correct values.
1
At the
Calibrate
level, press the key:
The
Calibrate
menu is displayed:
2
Use the
/
keys to select the sensor to be
calibrated, and press the key to confirm selection.
The pH calibration menu is displayed:
3
Use the
/
keys to select sensor calibration and
press the key to confirm selection.
The calibration type is displayed:
4
Use the key to edit the calibration type.
Use the
/
keys to select the required calibration
type and press the key to confirm selection.
Press the key to proceed to the next step.
5
Place the sensor into buffer 1 and press the
key to perform the calibration. The calibration
process screen is displayed.
On completion the result screen is displayed.
• If the calibration passes, the slope and offset values are
displayed.
• If the calibration fails, the failure reason is displayed.
Note
. The calibration can be canceled at any time during
the process by pressing the
Abort
key.
Calibrate
Back
Select
S1 : pH/Redox (ORP)
S2 : pH/Redox (ORP)
pH Buffers
Hold Outputs
Calibrate
Back
Select
Sensor Cal
Temperature Cal
Edit Cal
Sample Collection
Collection Complete
Restore Cal Defaults
Sensor Cal
Next
Edit
Calibration Type
1-Pt Auto
Calibrate
Abort
Continue
Immerse in Buffer 1
ABB
4.0 pH
Calibrate
Abort
pH 4.00 pH
Tmp 25.0 °C
Settling-Please Wait