82
AW T4 20 |
U N I V ER S A L 4 - W I R E , D UA L- I N P U T T R A N SM I T T ER | O I/AW T4 2 0 - EN R E V. B
…15 Calibration procedures
…Dissolved oxygen calibration
2-Point calibration
1
At the
Calibrate
level, press the key:
The
Calibrate
menu is displayed:
2
Use the
/
keys to select the sensor to be calibrated,
and press the key to confirm selection.
The RDO calibration menu is displayed:
3
Press the key to confirm selection.
The
Calibration Type
is displayed:
4
Use the key to edit the
Calibration Type
.
Use the
/
keys to select the required
Calibration Type
and press the key to confirm selection.
Press the and key to proceed to the next step.
5
Ensure the sensor is subjected to 100 % saturation and
press the key to start the calibration process.
The calibration process screen is displayed:
• If the calibration fails the result screen is displayed with
the reason for failure.
• If the calibration passes the procedure moves
automatically to the high buffer calibration.
6
Ensure the sensor is subjected to 0 % saturation and
press the key to start the calibration process.
The calibration process screen is displayed:
On completion the result screen is displayed.
• If the calibration passes the new settings are displayed.
• If the calibration fails the reason for failure is displayed.
Note
. The calibration can be canceled at any time during the
process by pressing the
Abort
key ( ).
Calibrate
Back
Select
S1 : RDO
S2 : RDO
Hold Outputs
Calibrate
Back
Select
Sensor Calibration
Sensor Calibration
Calibration Type
2-Point
Next
Edit
Calibrate
Start Span Cal?
Place in 100% Sat
Abort
Continue
Calibrate
PV 98.26 %
TMP 24.22 °C
Settling - Please Wait
Abort
Calibrate
Start Zero Cal?
Place in 0% Sat
Abort
Continue
Calibrate
PV 2.06 %
TMP 24.84 °C
Settling - Please Wait
Abort