7 CALIBRATION
64
IM/AX4CO Issue 11
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
Note.
The upper 7-segment display shows the measured conductivity. Once the signal
is within range the lower 7-segment display shows the same value and
Calib
is
displayed to indicate that calibration is in progress.
Resistance Span (2k0)
Set the cell simulator to 2k
Ω
.
The display advances automatically to the next step once a stable and valid value is
recorded.
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
Continued on next page.
A:Res.Span (2K0)
500.0
500.0
uS/cm
Calib
A:Res.Zero (O/C)
0
0
uS/cm
Calib
A:Res.Span(20R)
A:Res.Zero (O/C)
0.0
0.0
uS/cm
Calib