Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
7 CALIBRATION
62
IM/AX4CO Issue 11
7.3 Factory Settings
Fig. 7.1 Analyzer Terminal Links and Decade Resistance Box Connections
Fig. 7.2 Overall Factory Settings Chart
Note.
The Option Board page is displayed only if the option board is fitted.
Terminal link
Conductivity
Cell Simulator
Temperature
Simulator
Earth
B10
B11 B12
B13
B14
B9
Sensor A Terminal Numbers
Sensor B Terminal Numbers
B2
B3
B4
B5
B6
B1
FACTORY SETTINGS
Factory Set Code
Cal. Sensor A
A:Res.Zero (O/C)
A:Res.Span (2K0)
A:Res.Zero (O/C)
A:Res.Span(20R)
A: Self Checking
A:T.Zero (100R)
A:T.Span (150R)
Cal. Sensor B
B:Res.Zero (O/C)
B:Res.Span (2K0)
B:Res.Zero (O/C)
B:Res.Span(20R)
B: Self Checking
B:T.Zero (100R)
B:T.Span (150R)
Cal. Output 1
O1: Adjust 4mA
O1: Adjust 20mA
To
OPERATING PAGE
A:T.Zero (1K0) B:T.Zero (1K0)
A:T.Span (1K5) B:T.Span (1K5)
Dual conductivity only
Key
Alter Fact.Code
Cal. Output 3
Cal. Output 4
O3: Adjust 4mA
O4: Adjust 4mA
O3: Adjust 20mA
O4: Adjust 20mA
Cal. Output 2
O2: Adjust 4mA
O2: Adjust 20mA
Use the Sidescroll Key to scroll through the Pages within each Menu
Use the Menu
Key to scroll
through
the Menus
Use the Downscroll
Key to scroll through
the Frames
within each Page
Section 7.3, Page 63
Option Board
Available only if the option board is fitted
and
analog features enabled – see page 67
Calibrate I/P ?
Calibrate I/P ?