Aztec 600 | Piston assembly | Single- and multi-stream colorimetric analyzers |
SI/ANAINST/001-EN
5
4.2 Initiating a manual piston clean / reset
To initiate a manual clean / reset:
1.
At the
Indicator View
screen, press the
key to display the
Operate
menu and use the
key to highlight
Diagnostics
:
2.
Press the
key to select
Diagnostics
sub-menu options
and use the
and
keys to highlight
Cell Diagnostics
:
3.
Press the
key to display
Cell Diagnostics
sub-menu
options:
4.
Use the
and
keys to navigate between fields and
enter
data in the first 3 fields as shown below:
5.
Use the
key to highlight the
button press
to initiate
the clean.
6.
When the clean is complete,
the
and
keys to navigate
to the
Operation
field and select
Reset Pump
.
7.
The piston can now be removed from the instrument and
returned to the factory for repair – refer to Section 4.3,
page 6.
IMPORTANT (NOTE)
Perform this procedure to clean the piston of any
solution / chemicals and reset it before returning it to
the factory for repair.
IMPORTANT (NOTE)
If this procedure is selected while the analyzer is
operating, the following warning is displayed with
the
Continue
option active:
To exit press the
key, to proceed, press the
key.
Diagnostics
Cell Diagnostics
Monitor Status
Fields
Selection / Data
Operation
Bring In
Volume 9
ml
Port
DI
Detector
Voltage
Read-only
(detector voltage [0 to 4095 mV] in real time).
LED Current
Read-only
(LED current in real time).
Fields
Selection / Data
Operation
Reset Pump