background image

4

2 ELECTRICAL INSTALLATION

2.1 Electrical Connections – Figs. 2.1 to 2.3

Warning.

 Before making any connections, ensure that the

power supply, any high voltage-operated control circuits

and high common mode voltages are switched off.

Note.

• Use flexible conduit for cable routing 

(signal and power).

• Always route signal leads and power cables separately,

preferably in earthed metal conduit.

• Screened cable must be used for signal inputs and relay

connections. Connect the screen to the ground stud.

• The terminal blocks can be removed from the main p.c.b.

when making connections. Before removing any module

note its position – see Fig. 2.1.

Information.

 Use cable appropriate for the load currents.

The terminals accept cables up 12AWG for power supply

connections and 14AWG for all other connections.

Unplug
Module

Release
Clip

Remove Terminal
Block Assembly 

1

2

3

Fig. 2.2 Removing Terminal Block Assembly

DAN

GE

R

LIV

E

TE

RM

INA

LS

Remove securing screw

Remove
Cover

1

2

Fig. 2.1 Removing High Voltage

 Digital Interface Card Cover

Summary of Contents for C1900 Series

Page 1: ...ch for or click on C1900 Circular chart recorder Datasheet DS C1900R EN C1900 Circular chart recorders and recorder controllers Installation Guide IM C1900 INS C1900 Circular chart recorders Operating...

Page 2: ...roval of the Technical Publications Department Health and Safety To ensure that our products are safe and without risk to health the following points must be noted The relevant sections of these instr...

Page 3: ...This Supplementary Manual provides an overview of the use of the C1900 Circular Chart Recorder on a Dairy Farm and describes the electrical connections and special alarm feature designed for this app...

Page 4: ...arning light also located in the barn is illuminated if milk temperature and or time limits are exceeded An additional set of indicator and warning lights can be mounted in a more convenient location...

Page 5: ...Red White Red White C1900 Milking System Power Supply A B E D C C1900 Connections A Analog input connection from tank RTD sensor B Analog input connection from return flowline RTD sensor C Digital in...

Page 6: ...used for signal inputs and relay connections Connect the screen to the ground stud The terminal blocks can be removed from the main p c b when making connections Before removing any module note its po...

Page 7: ...y Connections within Recorder 230 V AC Connections 115 V AC Connections 24 V AC DC Connections Terminal Block 2 Channel 2 Green High Voltage Digital Interface Card See Fig 2 1 for cover removal detail...

Page 8: ...2 1 dEGF Lower the chart clamp and press down firmly to ensure that the 2 locating pins pierce the paper 1 2 3 4 5 6 3 1 Fitting the Chart Fig 3 1 75 0 Raise pens Remove cap Fit new pen capsule ensuri...

Page 9: ...enabled and delay time has expired Alarm Enable signal is switched OFF by activation of refrigeration system Alarm is disabled immediately Alarm de activates 1 2 3 4 5 6 7 Note Refer to Programming G...

Page 10: ...A s e Y e l b a n E t f i L n e P d e l b a n E s m r a l A e p y T t n e m e g d e l w o n k c A e n o N e c r u o S k c A l a b o l G e n o N 1 A m r a l A 1 B m r a l A s m r a l A r e h t O l l A...

Page 11: ...Sales Software Service...

Page 12: ...ll rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden w...

Reviews: