13
Electromagnetic Flowmeter
COPA-XE / MAG-XE
Submenu/Parameter
Entry Type
Comments
from table/numeric
Exit the submenu
ml/s, ml/min, ml/h, Ml/h, Ml/min, Ml/day, lbs/s,lbs/min, lbs/h,
uton/min, uton/h, uton/day, l/s, l/min,l/h, hl/s, hl/min, hl/h,
m3/s, m3/min, m3/h, igps, igpm, igph, mgd, gpm, gph,
bbl/s, bbl/min, bbl/h, bls/day, bls/min, bls/h, kg/s, kg/min, kg/h,
t/s, t/min, t/h, g/s, g/min, g/h, kgal/s, gkal/min, kgal/h
ml, l, hl, m3, igal, gal, mgal, bbl, bls, kg, t, g, Ml, lb, uton, kgal
If the desire units are not included in the table, it is possible to
configure a user defined unit, based on liters. The value shown,
3785.41, is the conversion factor for kgal (Factory setting).
Four character name for the user configured unit.
Programmable unit for mass (with density) or volumetric flow
(without density)
from table/numeric
Exit the submenu
All detected errors (Errors 0-9, A, B, C) are stored.
The error register can be cleared by pressing ENTER.
Limit alarm, range 0 - 130 % of the flow range setting.
Can be set in steps of 1 %, switching hysteresis 1 %.
Limit alarm, range 0 - 130 % of the flow range setting.
Can be set in steps of 1 %, switching hysteresis 1 %.
from table
Contact output, Terminal P7/G2 selections (for Profibus terminals
Ux/P7): General alarm
)
, empty pipe
1)
, F/R signal, no function,
MAX-Alarm
1)
, MIN-Alarm
1)
,MAX/MIN-Alarm
1)
1) Contact input can be configured to “Open or Close”.
Contact input, Terminal X1/G2 selections: External zero return,
external totalizer reset, external totalizer stop, no function. For
HART-Protocol the external totalizer stop function is not available.
The contact input is not available for Profibus.
Submenu
Units
C/CE
ENTER
Units
Qmax
l/s
Units Totalizer
m3
Units Factor
3785.41 Liter
Unit Name
kgal
/s
/min
/h
Prog. Units
w/o Density
Submenu
Alarm
C/CE
ENTER
Error Register
0 ... 3 ...
Max-Alarm
130 %
Min-Alarm
10 %
Submenu
Prog. In-/Output
ENTER
Terminals P7/G2
General Alarm
Terminals X1/G2
Ext. Zero Return
Summary of Contents for COPA-XE
Page 4: ......