Introduction 

Based on field experience, this guide is intended 
to help commission the ABB CoriolisMaster 
flowmeter. 

 

 

Additional Information 

Additional documentation on CoriolisMaster FCB100, 
FCB400 is available for download free of charge at 
www.abb.com/flow 
 
Alternatively simply scan this code: 
 

 

 

— 
We reserve the right to make technical changes or modify the contents of this document without 
prior notice. With regard to purchase orders, the agreed particulars shall prevail.  
ABB does not accept any responsibility whatsoever for potential errors or possible lack of 
information in this document.  
 
We reserve all rights in this document and in the subject matter and illustrations contained therein. 
Any reproduction, disclosure to third parties or utilization of its contents – in whole or in parts – is 
forbidden without prior written consent of ABB. 
 
© ABB 2018 

3KXF411019R0601 

 

 
 

Summary of Contents for CoriolisMaster FCB100

Page 1: ...ment made easy FCB100 FCB400 Introduction Based on field experience this guide is intended to help commission the ABB CoriolisMaster flowmeter Additional Information Additional documentation on Coriol...

Page 2: ...CB100 7 In Situ proving with a flow computer 7 Proving connection when the meter has been removed for the line 7 CoriolisMaster FCB400 8 In Situ proving with a flow computer 8 Proving connection when...

Page 3: ...in 4 bbl h 168 ugal h DN 25 1 in 15 bbl h 630 ugal h DN 50 2 in 45 bbl h 1890 ugal h DN 80 3 in 115 bbl h 4830 ugal h DN 100 4 in 220 bbl h 9240 ugal h DN 150 6 in 270 bbl h 11340 ugal h The proving m...

Page 4: ...ponse and high repeatability mode utilizing the following configuration settings 1 Damping Mass 0 1 s 2 Damping Density 0 1 s 3 Volumeflow QV set to your required volume flow unit e g bbl d 4 Volume T...

Page 5: ...imum nor 90 of the maximum rated input frequency of the tertiary devices receiving the signal Upper frequency calculation example The upper frequency should be calculated in the following way Upper fr...

Page 6: ...the resistors can be integrated in the connection area of the Coriolis sensor connection box This is the case if the sensor needs to be disassembled out of the pipe for proving 1 Input channel for fl...

Page 7: ...ing In the case that the flow computer panel provide a connector for the prover 1 Clamping connection 2 Cable gland for prover connection Threshold at prover and flow computer should be 3 5 V 0 5 V Fi...

Page 8: ...d be 3 5 V 0 5 V Figure 11 Connection for proving Proving connection when the meter has been removed from the line R1 1 k 0 5 W between terminal 32 and 41 R2 1 k 0 5 W between terminal Uco and 41 Jump...

Page 9: ...sMaster FCB100 or FCB400 How to connect to a flow computer and to a prover by using two outputs R1 1 k 0 5 W between GND and input pulses R2 2 k 0 5 W between 24 V DC and input pulses Jumper wire betw...

Page 10: ...lity of the density meter factor DMF should be 0 05 or better between two pycnometer measurements in series 7 Of course a check of the detector switches of the prover and the certified measuring secti...

Page 11: ...027 9 0 10 0 027 10 0 12 0 027 11 0 13 0 027 12 0 14 0 027 13 0 15 0 027 14 0 16 0 027 15 0 17 0 027 16 0 18 0 027 17 0 19 0 027 18 0 20 0 027 19 0 21 0 027 20 0 22 0 027 Sets of runs can also be ave...

Page 12: ...than 10000 per proving cycle Make sure that the pulse per unit match between the flowmeter and the flow computer Figure 14 Digital output configuration as pulse output hange from one to two columns No...

Page 13: ...FCB100 FCB400 COMMISSIONING AND PROVING PG CORIOLIS PROVING EN REV A 13 Trademarks Modbus is a registered trademark of the Modbus Organization...

Page 14: ...14 FCB100 FCB400 COMMISSIONING AND PROVING PG CORIOLIS PROVING EN REV A Notes...

Page 15: ...FCB100 FCB400 COMMISSIONING AND PROVING PG CORIOLIS PROVING EN REV A 15 Notes...

Page 16: ...83 PG CORIOLIS PROVING EN Rev A 02 2019 We reserve the right to make technical changes or modify the contents of this document without prior notice With regard to purchase orders the agreed particular...

Reviews: