29
4
CONFIGURATION MODE
…4.4
Ranges and Passwords (Level 4)
Volume Constant – Fig. 4.10
This constant can be used to calculate the volume of
liquid in a tank from the measured level
[0.0 to 999.9]
Set Up Password
The password enables access to the set up level
(Level 2).
[0 to 9999] Default/factory set = 0
Configuration Password
The password enables access to the configuration
levels (Levels 3 and 4). The Configuration password
overides the setup password.
[0 to 9999] Default/factory set = 0
MODBUS Address
This frame sets the MODBUS address.
[1 to 99]
1.0
VOLCON
0
S-PASS
0
C-PASS
1
Addr.
•1
•1 Only available if the appropriate option board is fitted.
Fig. 4.10 Volume Constant Calculations
d
L
Cylinder:
Volume =
π
x ( d/2)
2
x L
Volume
Constant
=
π
x ( d/2)
2
L
b
a
Cube:
Volume =
a x b x L
Volume
Constant
=
a x b