… 10 Configuration
Magnos206 – Change measuring range
limits
Menu Path
‘
Setup /
Measurement Ranges’
The measuring range limits can also be set in the Configurator,
see
Measuring range configuration with ECT
Note
• Measuring ranges may not be specified within ignition limits.
• After changing the measuring range limits, the calibration of
the relevant measuring range needs to be checked.
Measurement range limits
The
lower range value
and the
upper range value
must be within
the physical range.
Setting range
The measuring range can be freely set within the limits of the
physical measuring range.
• The measuring range set at the factory can be found on the
name plate or device data sheet.
• The minimum measuring range is 0 to 2 Vol.-% O
2
Measuring range with suppressed zero point (‘suppressed
measuring range’)
• The minimum measuring span of the suppressed measuring
range is 2 Vol.-% O
2
.
• The measuring range set at the factory can be found on the
name plate or device data sheet.
• If the analyzer has been ordered with the suppressed
measuring range, a pressure sensor is installed at the
factory.
Physical measuring range
The analyzer has a physical measuring range. The limits of this
measuring range are 0 Vol.-% O
2
or 100 Vol.-% O
2
.
Magnos28 – Change measuring range
limits
Menu Path
‘
Setup /
Measurement Ranges’
The measuring range limits can also be set in the Configurator,
see
Measuring range configuration with ECT
on page 142.
Note
• Measuring ranges may not be specified within ignition limits.
• After changing the measuring range limits, the calibration of
the relevant measuring range needs to be checked.
Measurement range limits
The
lower range value
and the
upper range value
must be within
the physical range.
Setting range
The measuring range can be freely set within the limits of the
physical measuring range.
• The measuring range set at the factory can be found on the
name plate or device data sheet.
• The minimum measuring range is 0 to 0.5 Vol.- % O
2
.
Measuring range with suppressed zero point (‘suppressed
measuring range’)
The minimum measuring span of the suppressed measuring
range is 0.5 Vol.-% O
2
• The measuring range set at the factory can be found on the
name plate or device data sheet.
• If the analyzer has been ordered with the suppressed
measuring range, a pressure sensor is installed at the
factory.
Physical measuring range
The analyzer has a physical measuring range.
The limits of this measuring range are 0 Vol.-% O
2
or 100 Vol.-%
O
2
.