124
EL3000
CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D
… 9 Calibration
… Manual calibration: calibration method
Calibration method (manual)
Analyzer
Calibration method
Remark
Uras26
Zero point / final point calibration (test gas)
Cannot be changed.
Limas23
Zero point / final point calibration (test gas)
Cannot be changed.
Magnos206
Zero point / final point calibration (test gas)
or
Substitute gas or single point calibration
If a substitute gas component is set up in the Magnos206 analyzer, the calibration
method is factory-set to substitute gas calibration, otherwise the zero point / end-
point calibration (test gas) is set.
Alternatively, single-point calibration can be selected.
See also page 130.
Magnos28
Zero point / final point calibration (test gas)
or
Substitute gas or single point calibration
If a substitute gas component is set up in the Magnos28 analyzer, the calibration
method is factory-set to substitute gas calibration, otherwise the zero point / end-
point calibration (test gas) is set.
Alternatively, single-point calibration can be selected.
See also page 131.
Magnos27
Zero point / final point calibration (test gas)
or
Substitute gas calibration
If a substitute gas component is set up in the Magnos27 analyzer, the calibration
method is factory-set to substitute gas calibration, otherwise the zero point / end-
point calibration (test gas) is set.
See also page 133.
Caldos27
Zero point / final point calibration (test gas)
or
Single-point calibration
Factory set
or
With standard gas
Fidas24
Substitute gas calibration
Factory set
Oxygen sensor
Zero point / end-point calibration
Cannot be changed, the oxygen sensor is only calibrated at the end point.
See also page 140.