EL3000
CONTINUOUS GAS ANALYZERS | OI/EL3000-EN REV. D
129
Test gases – Limas23
Analyzer(s)
Test gas for the zero calibration
Test gas for the end-point calibration
Limas23
with calibration cells
(automatic calibration)
N
2
or air or UV sample component-free gas
Calibration cells or test gas for each sample
component
Limas23
without calibration cells
(automatic calibration)
N
2
or air or UV sample component-free gas
Test gas for each sample component
Limas23
without calibration cells
(manual calibration)
N
2
or air or UV sample component-free gas
Test gas for each sample component
L Magnos206 / Magnos28
or
oxygen
sensor
with calibration cells
(automatic calibration, i.e. Magnos206 / Magnos28
with single-point calibration)
N
2
or oxygen and UV sample component-free gas
Either calibration cells and test gas for oxygen
detector or test gas for each sample component or
for each detector
L Magnos206 / Magnos28
or
oxygen
sensor
without calibration cells
(automatic calibration)
N
2
or oxygen and UV sample component-free gas
Test gas for each sample component or for each
detector
L Magnos206 / Magnos28
or
oxygen
sensor
without calibration cells
(manual calibration)
N
2
or oxygen and UV sample component-free gas
Test gas for each sample component or for each
detector
Dew point
The dew point of the test gases must be approximately equal to the dew point of the sample gas.