24
EL3000
CONTINUOUS GAS ANALYZERS | CI/EL3000-EN REV. B
Test gases
Analyzer
Test gas for zero-point calibration
and single-point calibration
Test gas for end-point calibration
Magnos206
Oxygen-free process gas
Process gas with known O
2
concen-
tration
Magnos206 with
suppressed measuring
range
Test gas with O
2
concentration near
the starting point of the measuring
range
Test gas with O
2
concentration near
the end point of the measuring range
Magnos206 with
single-point calibration
Test gas with O
2
concentration in an
existing measuring range or ambient
air. Same moisture content as pro-
cess gas.
–
Magnos206 with
substitute gas calibration
Oxygen-free process gas or substi-
tute gas (O
2
in N
2
)
Substitute gas, e.g. dried air
CAUTION
In order to avoid accumulations of explosive gas mixtures, do not use air
as a test gas for single-point calibration when measuring flammable gas-
es!
Dew point
The dew point of the test gases must be approximately the same as the dew
point of the sample gas.
NOTE
The notes for calibrating must be considered.
Pressure sensor
The pressure sensor is installed in the gas analyzer as an option. It is con-
nected to a connection port via an FPM tube.
With measurements in suppressed measuring ranges the connection of the
pressure sensor and the sample gas outlet are to be connected to each oth-
er via a T-joint and short lines.
Particular care should be taken that the gas discharge line is as short as
possible or – in the case of a greater length – has a sufficiently wide inside
diameter (at least 10 mm).
Gas connections
see sections
"Gas connections Magnos206 (Model EL3020)" (see page 69) and
"Gas connections Magnos206 (Model EL3040)" (see page 70)