70
EL3000
CONTINUOUS GAS ANALYZERS | CI/EL3000-EN REV. B
Gas connections for Magnos206 (Model EL3040)
Magnos206: Gas connections
1
Sample gas inlet
2
Sample gas outlet
3
not assigned
4
not assigned
11
Purge gas inlet for case
12
Purge gas outlet for case
13
Pressure sensor
Design:
⅛
NPT female thread (stainless steel 1.4305/SAE 303)
Connection of flexible tubes: straight screwed fittings (PP) with nozzles for flexible tubes with
inside diameter = 4 mm (supplied)
Connection of pipelines: Threaded connections (not supplied)
Note:
The pressure sensor is installed as an option. It is connected to the port
13
via an FPM hose. The
pressure sensor connection and the sample gas outlet must be connected to each other via a
T-piece and short lines to ensure a precise pressure correction. The lines must be as short as
possible or if this is not practicable, the line must have a sufficiently wide internal diameter
(at least 10 mm) so that the flow effect is minimized. If the pressure sensor connection is not
connected to the sample gas outlet, the pressure sensor and the sample gas outlet must be on
the same pressure level.