7.3.6 Diagnostics
Used to perform a loop test (output verification) and indicates
the diagnostic status (actual / simulated).
Diagnostics Parameters
Parameter
Description
Tx
Type
PV cal alarm lim
All
Slope alarm lim high
100 % to 150 %. Default = 110 %.
Slope alarm lim lo
40 % to 100 %. Default = 60 %.
Offset alarm Lim +/–
0 to ±10000 mV. Default = ±180 mV.
Loop Test
The current output of the transmitter can be adjusted between 3.8 and
21.5 mA.
This feature is useful to verify that the transmitter output matches the current
in the control room.
Enabled when
HART Poll Addr
=
0
– see page 69.
Diagnosis Status
Used to determine if diagnostics simulation is active.
When display reads
Actual
, real-time diagnostics from the transmitter are
active.