pH / Redox (ORP) transmitter
7 Operator Pages and Menus
54
OI/APA592–EN Rev. E
7.3 Parameters
7.3.1 Easy Setup
Users with
Advanced
level access have read / write privileges
to the parameters listed.
Users with
Standard
and
Read Only
access have read-only
privileges to parameters at
Easy Setup
level.
Easy Setup Parameters
Parameter
Description
Tx
Type
Analyzer Type
Selects the measurement type:
pH
ORP
pION
Ion Conc
All
pH Sensor Type
Selects the sensor type used for measurement:
Glass
Antimony
Custom
pH
Isopotential Pt.
(Isopotential Point)
Adjustable range –20 to 20 pH. Default = 7 pH.
Enabled when
pH Sensor Type
=
Custom
.
pH
Asymmetric Pt
.
(Asymmetric Potential)
Adjustable range –2000 to 2000 mV. Default = 0 mV
Enabled when
pH Sensor Type
=
Custom
.
PH
Valence
Ion valence selection: -3, -2, -1, 1, 2, or 3.
The valence determines the millivolt change per decade of concentration.
Ion
Conc
Magnitudes
The number of magnitudes ranging from 1 to 3 that defines the ion
concentration output.
Ion
Conc
End Mag
The ion concentration state functions by associating an end millivolt value to
an end magnitude value.
The end magnitude value can be set to 10, 100 or 1000.
Ion
Conc
End mV
The end millivolt value associated with the end magnitude value.
Ion
Conc
PV Units
Sets the process value units (%, ppm, µg/l, ppb, mg/l).
Ion
Conc
Temp. unit
(Temperature Units)
Sets the temperature unit (°C or °F) applied to all transmitter temperature
parameters and values.
All
1
Menu
Exit
Select
Easy Setup