Endura AZ series integral and remote transmitter
Combustion oxygen monitor
5 Programming
54
IM/AZ20E–EN Rev. F
5.4.7 Input/Output
Used to assign relays (1 and 2), digital I/O (1 and 2) and
current outputs (1 and 2).
Menus displayed at this level are dependent on the system
configuration / options fitted.
Parameter
Comment / [Range]
Default
Relay 1 (2)
Relay Assignment
Selects the alarm and diagnostic states required to
activate the relay.
Each
Relay Assignment
parameter can be
Assigned
or
Not Assigned
.
Note
.
Assigned
parameters are OR'd together.
Not
Assigned
(all parameters)
Alarm 1 (2, 3, 4)
See Section 5.4.6, page 53 for alarm types and
triggers.
Cal. In Progress
Relay is activated when a calibration is in progress.
Cal. Failed
Relay is activated if a calibration fails.
Gas 1 Not Present
Relay is activated if test gas 1 is not detected.
Gas 2 Not Present
Relay is activated if test gas 2 is not detected.
Test Gas 1 Valve
Relay is activated if test gas 1 valve is open.
Test Gas 2 Valve
Relay is activated if test gas 2 valve is open.
Diag. – Failure
Activates the relay if a
Failure
diagnostic status is
generated – see Section 8.1, page 78.
Diag. – Off Spec.
Activates the relay if an
Out of Specification
diagnostic status is generated – see Section 8.1,
page 78.
Diag. – Maint. Reqd.
Activates the relay if a
Maintenance Required
diagnostic status is generated – see Section 8.1,
page 78.
Diag. – Chk.
Function
Activates the relay if a
Check Function
diagnostic
status is generated – see Section 8.1, page 78.
Table 5.12 Input/Output Menu
7
Menu
Exit
Select
Input/Output