Endura AZ series integral and remote transmitter
Combustion oxygen monitor
8 Troubleshooting
78
IM/AZ20E–EN Rev. F
8 Troubleshooting
8.1 Diagnostics Classification Codes
Diagnostic messages and icons conforming to the NAMUR NE107 classification code are used to define
information during operation and data entry – a typical diagnostic message is shown in Fig. 8.1:
Fig. 8.1 Example Diagnostic Message and Components
Item
Description
Item
Description
1
NAMUR status Icon (see
3
):
Failure
Out of Specification
Maintenance
Check
Function
4
Diagnostic Priority:
Failure
070 to 100
Out of Specification
048 to 068
Maintenance
022 to 042
Check Function
002 to 020
2
Group Name:
Electronics
Process
Configuration
Operation
5
Diagnostic Number (Internal Use Only):
Minimum 000
Maximum
047
3
NAMUR Classification Code:
C Check
Function
F Failure
M Maintenance
S Out of Specification
6
Diagnostic Text:
a unique message associated with a
diagnostic message
Table 8.1 Diagnostic Message Components
S070.003
Process
- - F070. 003 --
TEMP. SENSOR
OUT OF LIMITS
Check operating
conditions
Back
Exit