Endura AZ20 series probe
Combustion oxygen monitor
3 Mechanical Installation
12
IM/AZ20P–EN Rev. G
3.7 Endura AZ20 Probe Flanges – All Probe Lengths
Dimensions in mm (in).
Note.
These flanges are not pressure rated.
Flange Type
A
B
C
(Ø)
D
(PCD)
4-Hole Flanges
ANSI 2 in 150
152.4
(6.00)
12
(0.47)
19
(0.75)
120.6
(4.75)
ANSI 2.5 in 150
177.8
(7.00)
12
(0.47)
19
(0.75)
139.7
(5.50)
ANSI 3 in 150
190.5
(7.50)
12
(0.47)
19
(0.75)
152.4
(6.00)
DIN 65 PN16
185
(7.28)
12
(0.47)
18
(0.70)
145
(5.70)
JIS 65 5K
155
(6.10)
12
(0.47)
15
(0.59)
130
(5.12)
JIS 80 5K
180
(7.08)
12
(0.47)
19
(0.75)
145
(5.71)
Table 3.1 4-Hole Probe Flange Types and Dimensions
A
B
C
D
Flange Type
A
B
C
(Ø)
D
(PCD)
6-Hole Flanges
ABB Standard
(0.5 m [1.64 ft.]
probes only)
101
(3.97)
6 (0.24)
7.3
(0.29)
80
(3.15)
ABB Standard
165
(6.50)
12
(0.47)
12.5
(0.50)
140
(5.51)
Table 3.2 6-Hole Probe Flange Types and Dimensions
A
B
D
C
Flange Type
A
B
C
(Ø)
D
(PCD)
8-Hole Flanges
ANSI 4 in 150
228.6
(9.0)
12
(0.47)
19
(0.75)
190.5
(7.50)
DIN 80 PN16
200
(7.87)
12
(0.47)
18
(0.70)
160
(6.30)
DIN 100 PN16
220
(8.66)
12
(0.47)
18
(0.70)
180
(7.08)
JIS 100 5K
200
(7.87)
12
(0.47)
19
(0.75)
165
(6.50)
Table 3.3 8-Hole Probe Flange Types and Dimensions
A
B
D
C
Summary of Contents for Endura AZ20 series