Combustion Oxygen Monitor
Endura AZ20 Series Probe
5 Dismantling and Reassembly – Probe
26
IM/AZ20M–EN
Rev. B
5.9.2 Dismantling the Heater Assembly
Referring to Fig. 5.11:
1. Remove the red tube
A
and pull the sleeved heater wires
B
through the internal structure mounting plate / end
plate
C
.
2. Remove and discard the wire
D
holding the 2 thermal
barrier sections
E
together and retain the barrier
sections for re-assembly.
3. Fold the insulator retaining tabs
F
back just enough to
free the insulators along the complete length of the inner
assembly.
4. Remove each insulator
G
by sliding it over the heater
wires.
5. Release the heater assembly
H
from the inner assembly
support column(s)
I
by folding the 4 tabs
J
on the
heater assembly end plate back.
6. Separate the heater assembly
H
from the inner support
sections
I
.
7. Discard the heater assembly
H
and insulators
G
.
8. Proceed to Section 5.9.3, page 27 to assemble the new
heater assembly.
Fig. 5.11 Dismantling the Existing Heater Assembly
G
D
I
E
B
C
A
F
J
H
*
*0.5 m (1.64 ft.) Probe shown,
1.0 m (3.28 ft.) probes and longer have multiple columns