Combustion Oxygen Monitor
Endura AZ20 Series Probe
Appendix B – Probe Electrical Connections
48
IM/AZ20M–EN
Rev. B
Appendix B – Probe Electrical Connections
B.1 AZ20 Probe Transmitter Cable Connections
B.2 AZ20/ZFG2 Replacement Probe
Transmitter Cable Connections
Note.
Non-AutoCal probes are not fitted with the 6-way
AutoCal terminal block or pressure switch / solenoid valve
option.
Terminal / Cable
Color
Tag ID
Tx Connection
Violet
ACJC
Pt1000 Cold
Junction Compensation
Grey
ACJC
Pt1000 Cold
Junction Compensation
Red
Cell +
Oxygen Input (+ve)
Black
Cell –
Oxygen Input (–ve)
Green
T / C+
Thermocouple (+ve)
White
T / C –
Thermocouple (–ve)
Light Yellow
SCN
Screen
Brown
H
Heater
Blue
H
Heater
Table B.1 Probe Transmitter Cable Connections
Terminal / Cable
Color
Tag ID
Tx Connection
Violet
Not Used
Grey
Red
Cell +
Oxygen Input (+ve)
Black
Cell –
Oxygen Input (–ve)
Green
T / C+
Thermocouple (+ve)
White
T / C –
Thermocouple (–ve)
Light Yellow
SCN
Screen
Brown
H
Heater
Blue
H
Heater
Table B.2 Probe Transmitter Cable Connections
Fig. B.1 Probe Transmitter Cable Connections
Transmitter terminal
plug connected
to transmitter
terminal block
*Violet and grey cables not used on
AZ20/ZFG2 replacement probes.
*