Parameterization
D184B097U02
FV4000-VT4/VR4 / FS4000-ST4/SR4
93
8.7.5 Standard
factor
Under constant operating conditions (constant pressure and temperature), the standard factor
can be entered here.
The standard factor is defined as the ratio between standard flow and operating flow:
Standard factor =
)
T
273
(
273
bar
013
,
1
)
p
bar
013
,
1
(
Qv
Qn
+
×
+
=
Qn Standard flow
Qv Operating
flow
P
Operating pressure [bar ü]
T Temperature
[°C]
ρ
V Operating
density
ρ
N Standard
density
Since the mass flow is constant, the following formula applies:
N
V
Qv
Qn
ρ
ρ
=
8.7.6 Operating
density
Please refer to the "AP-Calc" product selection and dimensioning program, which ABB provides
free of charge.