Parameterization
92
FV4000-VT4/VR4 / FS4000-ST4/SR4
D184B097U02
8.7.4 Error
3/9
Error detection for errors "3" and "9" can be activated/deactivated in the "Error 3/9" menu. The
following settings are possible:
Configuration
Function
"Errors 3+9 off"
(factory setting)
If the set measuring range is overshot by more than 3.125 %, the current
output will maintain the 20.5 mA it will have reached by this point until the
flow returns to a level below the upper range value set for the measuring
range.
"Both on"
If the set measuring range is overshot by more than 3.125 %, the current
output will switch to the set fault current (factory setting 22.4 mA). If
QmaxDN is overshot by more than 20 %, error "9" will also be output as a
warning before the device is overloaded.
"Error 3 off"
If the set measuring range is overshot by more than 3.125 %, the current
output will maintain the 20.5 mA it will have reached by this point. If
QmaxDN is overshot by more than 20 %, error "9" will also be output and
the current output will switch to the set fault current.
G00721
1
2
3
20 mA
4 mA
0 %
Qmin
120 % von
QmaxDN
20,5 mA
100 %
103,125 %
Fig. 47: Current output in the event of an alarm
1
Current output without errors "3" and "9",
output: 20.5 mA (NAMUR NE43)
2
Current output with errors "3" and "9", the
output switches to alarm status (21 ... 23
mA, configurable)
3
Current output with error "9", the output
switches to alarm status at 120 % of
QmaxDN (21 ... 23 mA, configurable)
Qmin Low flow
Summary of Contents for FS4000-SR4
Page 125: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 125 ...
Page 126: ...Appendix 126 FV4000 VT4 VR4 FS4000 ST4 SR4 D184B097U02 ...
Page 127: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 127 ...
Page 128: ...Appendix 128 FV4000 VT4 VR4 FS4000 ST4 SR4 D184B097U02 ...
Page 129: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 129 ...
Page 135: ......