Parameterization
94
FV4000-VT4/VR4 / FS4000-ST4/SR4
D184B097U02
8.8
Parameterization of gases, steams, and liquids
The selection of possible operating modes, the required parameters associated with them and
the additional menu items that become visible are summarized in the following tables.
Important
The selection of possible operating modes is dependent upon how the flowmeter is calibrated.
Operating mode Medium
Flow type
Calculation
Correction
parameter
Additional
menus that
become visible
Fluid Qv
Volume flow
-
-
-
Fluid Qm(D)
b
Qv
Qm
ρ
×
=
Reference density
constant
Unit of density
Reference density
Unit Qm
Fluid
1)
Qm (D, T)
( )
b
T
Qv
Qm
ρ
×
=
( )
(
)
(
)
2
T
T
1
T
0
b
b
β
×
−
+
×
ρ
=
ρ
Reference density
ρ
b
Reference
temperature T
0
Temperature
measured T
b
Density compensating
coefficient ß2
Unit of density
Reference density
Reference
temperature
Unit Qm
D. comp. coeffi.
Fluid
1)
Qm (V, T)
Fluid
Mass flow
b
Qn
Qm
ρ
×
=
(
)
(
)
1
T
T
1
Qv
Qn
0
b
β
×
−
+
=
Vol.expansion
coefficient [%/K]ß1
Reference
temperature T
0
Temperature
measured T
b
Reference density
ρ
b
Unit of density
Reference density
Reference
temperature
Vol._expansion
Unit Qm
Qm Mass
flow
Qv
Operating flow
Qn
Standard flow
Pbtr Reference
pressure
β
1
Volume expansion coefficient
β
2
Density compensating coefficient
ρ
0
Standard density
ρ
b
Reference density
1) These operating modes can only be selected if the flowmeter is fitted with a temperature measurement device.
Summary of Contents for FS4000-SR4
Page 125: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 125 ...
Page 126: ...Appendix 126 FV4000 VT4 VR4 FS4000 ST4 SR4 D184B097U02 ...
Page 127: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 127 ...
Page 128: ...Appendix 128 FV4000 VT4 VR4 FS4000 ST4 SR4 D184B097U02 ...
Page 129: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 129 ...
Page 135: ......