Error
messages
D184B097U02
FV4000-VT4/VR4 / FS4000-ST4/SR4
97
9.2
Error description
Error no. Plain text message
Priority
Description
Possible cause
Troubleshooting
Steam
temperature
< 55 °C (131 °F)
Increase steam
temperature
0 Steam
calculation 7
Mass flow for
saturated steam
calculated incorrectly
Steam
temperature
> 370 °C (698 °F)
Reduce steam
temperature
1 Front
End 1
Problem affecting
preamplifier board
-
Replace transmitter
plug-in module /
Contact ABB Service
2 Not
assigned -
-
-
-
Measuring range
too small
Increase "Qmax!"
measuring range
3
Flow > 3 %
2
The value set in
Qmax is being
exceeded by more
than 3 %
Flow too large
Reduce flow
4 Not
assigned -
-
-
-
5 M
database 0
Main database
damaged beyond
repair, loss of
transmitter's internal
database
Reduce flow
Replace meter and
switch back on, replace
transmitter plug-in
module
if necessary, contact
ABB Service
6
Meter damaged
beyond repair
1
Flowmeter damaged
beyond repair, values
displayed are invalid
-
Reprogram the meter
PT100 faulty
Replace sensor
7 Temperature 7
Problem affecting
temperature
measurement
Wiring errors
between sensor
and transmitter
affecting VR / SR
models
Check wiring
8 Not
assigned -
-
-
-
9
Qv > 120 % of
QmaxDN
2
Maximum possible
measuring range
(QmaxDN) overshot
Flow too large
(software only)
level < A30
Reduce flow
B B
base 0
Backup database
damaged beyond
repair, loss of
external database
(sensor board)
External database
faulty
Switch the meter off
and back on again,
sensor board might be
faulty, contact ABB
Service
Summary of Contents for FV4000-VT4
Page 125: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 125 ...
Page 126: ...Appendix 126 FV4000 VT4 VR4 FS4000 ST4 SR4 D184B097U02 ...
Page 127: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 127 ...
Page 128: ...Appendix 128 FV4000 VT4 VR4 FS4000 ST4 SR4 D184B097U02 ...
Page 129: ...Appendix D184B097U02 FV4000 VT4 VR4 FS4000 ST4 SR4 129 ...
Page 135: ......