16 OI/FEP610/FEH610-EN Rev. B |
ProcessMaster FEP610, HygienicMaster FEH610
5.1.10
Free inlet or outlet
Fig. 16: Free inlet or outlet
A
Do not install the flowmeter at the highest point or in the
draining off side of the pipeline, flowmeter runs empty, air
bubbles can form.
B
Provide for a siphon fluid intake for free inlets or outlets so
that the pipeline is always ful.
5.1.11
Strongly contaminated measuring media
Fig. 17: Bypass connection
For strongly contaminated measuring media, a bypass
connection according to the figure is recommended so that
operation of the system can continue to run without
interruption during the mechanical cleaning.
5.1.12
Installation in the vicinity of pumps
Fig. 18: Vibration damping
1
Pump
2
Damping device
3
Shut-off device
Strong vibrations in the pipeline must be damped using flexible
damping devices.
The damping devices must be installed beyond the supported
flowmeter section and outside of the section between the
shut-off devices.
Do not connect flexible damping devices directly to the
flowmeter sensor.
G12010
A
B
G12011
G12012
1
2
3