ProcessMaster FEP610, HygienicMaster FEH610
| OI/FEP610/FEH610-EN Rev. B 33
6.4.1
Parameterization via the infrared service port
adapter
Configuration via the infrared service port adapter on the
device requires a PC / notebook and the FZA100 infrared
service port adapter.
All parameters can also be set using the HART DTM available
at www.abb.com/flow and the "ABB AssetVision" software.
Fig. 47: Infrared service port adapter on transmitter (example)
1
Infrared service port adapter
2
USB interface cable
3
PC / notebook running ABB AssetVision and HART DTM
1. Position the infrared service port adapter on the front plate
of the transmitter as shown
2. Insert USB interface cable into a free USB female
connector on the PC / notebook.
3. Switch on the device power supply.
4. Start ABB AssetVision and perform the parameterization
of the equipment.
Detailed information on operating the software is available in
the relevant operating instructions and the DTM online help.
6.4.2
Parameterization via the local operating interface
A PC / notebook and the USB interface cable are required to
configure the device via the device's local operating interface.
In conjunction with the HART-DTM and the software "ABB
AssetVision" available at www.abb.com/flow, all parameters
can also be set without the need for a local display.
Fig. 48: Connection to the local operating interface
1
Local operating interface
2
Programming plug
3
PC / notebook
4
USB interface cable
1. Open device terminal box.
2. Connect
programming plug to the local operating interface
of the device.
3. Insert USB interface cable into a free USB female
connector on the PC / notebook.
4. Switch on the device power supply.
5. Start ABB AssetVision and perform the parameterization
of the equipment.
Detailed information on operating the software is available in
the relevant operating instructions and the DTM online help.
G11911
RxD
TxD
3
2
1
G11625
1
2
3
4