ProcessMaster FEP610, HygienicMaster FEH610
| OI/FEP610/FEH610-EN Rev. B 51
7.5.3
Menu: Device Info
This menu is only used to display the device parameters. The parameters are displayed independently of the configured access
level, but cannot be changed.
Menu / parameter
Description
Device Info
...Sensor
Selection of submenu "
...Sensor
" using
.
...Transmitter
Selection of submenu "
...Transmitter
" using
.
Device Info /
...Sensor
QmaxDN Type
For information purposes only.
Meter Size
Nominal size of sensor.
Qm Max DN
The value specifies the maximum massflow rate a flow velocity of 10 m/s.
The value is set automatically via the selected nominal size multiplied with the density set.
Qv Max DN
The value specifies the maximum volume flowrate rate a flow velocity of 10 m/s.
The value is set automatically via the selected nominal size.
Mains Frequency
Mains frequency for the supply power.
Exitation Freq
Frequency used to operate the magnet coils of the flowmeter sensor.
Coil Current Regul.
For service information purposes only.
Set Point Curr.
Current used to operate the magnet coils of the flowmeter sensor.
DC feedback reg.
For service information purposes only.
Sensor ID
ID number of the sensor.
Sensor Serial No.
Serial number of the sensor.
Sensor Run Hours
Operating hours of the sensor.
...Se Calibration
Selection of submenu "
...Se Calibration
" using
.
Device Info / ...Sensor /
...Se Calibration
Se Span
Calibration value in the forward flow and reverse flow direction of the sensor.
Se Zero
First Cal. Date
Date of first calibration of sensor (calibration of new device).
Last Cal. Date
Date of last calibration of sensor.
Cal. Cert. No.
Identification (number) of the relevant calibration certificate.
First Cal. Location
Place of first calibration of the sensor.
Last Cal. Location
Place of last calibration of sensor.