ProcessMaster FEP610, HygienicMaster FEH610
| OI/FEP610/FEH610-EN Rev. B 67
Error no. / Range
Text on the LCD display
Cause
Remedy
S046.001 / Operation
Mass flowrate exceeds
limits.
Check flowrate and alarm
limits.
The mass flow is below or above the
configured limit values "Qm Massflow Min"
and "Qm Massflow Max".
Check the parameter settings in the "Process
Alarm / ...Alarm Limits" menu and adjust if
necessary.
Check mass flow.
S044.002 / Operation
Volume flowrate exceeds
limits.
Check flowrate and alarm
limits.
The volume flow is below or above the
configured limit values "Qv Volumeflow Min"
and "Qv Volumeflow Max".
Check the parameter settings in the "Process
Alarm / ...Alarm Limits" menu and adjust if
necessary.
Check volume flow rate.
S041.034 / Electronics
DC feedback regulation.
Check conditions of
application. Call service.
Multi-phase measuring media that produce a
very high level of noise. Stones or solids that
produce a very high level of noise.
Galvanic voltages at the measuring
electrodes.
Conductivity of measuring medium is not
evenly distributed (e.g., directly after injection
points).
Contact ABB Service.
S040.031 / Electronics
Coil impedance alarm.
Call service.
Coil Inductance changed, coil damaged, coil
isolation damaged, outer magnetic fields.
Contact ABB Service.
8.2.4
Maintenance
Error no. / Range
Text on the LCD display
Cause
Remedy
M038.010 / Electronics
Sensor memory defective.
Mem. or connect. defective.
Replace memory.
Defective NV-Memory module.
NV-Memory module is not plugged in.
Check if NV-Memory module is plugged in.
Call Service.
M037.015 / Electronics
NV chips defect on
Motherboard.
Defective MB. Replace MB.
Call Service.
Defective NV-Memory.
Exchange Motherboard.
Call Service.
M032.023 / Config.
Curr.Out 31 / 32 not
calibrated.
Call Service.
The Current Output 31 / 32 is not calibrated.
Call Service.
M028.008 / Config.
Display value is < 1600 h at
Qmax.
Change mass Unit or vol.
Unit for Totalizer.
Totalizer unit is too small.
Change the mass or volume totalizer unit.
M026.005 / Config.
Maintenance interval is
reached.
Perform maintenance.
Set "Preset Maint. cycle" to zero to
deactivate the maintenance timer.
Perform maintenance work.
Start new cycle.
M024.009 / Config.
Device not calibrated.
Call Service.
The device is not calibrated.
Call Service.