background image

ABB MEASUREMENT & ANALYTICS | USER MANUAL

User Manual | ICOS
GLA151 Series QC Portable Analyzers

Summary of Contents for ICOS GLA151 Series

Page 1: ... ABB MEASUREMENT ANALYTICS USER MANUAL User Manual ICOS GLA151 Series QC Portable Analyzers ...

Page 2: ...Performance Specifications 12 Ambient Humidity 12 Operating Temperature 12 Maximum Altitude 12 Power Requirements 12 Fuse Ratings 12 Cable Plugs and Voltage for EC Countries 12 External Dimensions 13 Weight 13 Cell Volume 13 Standard Components 14 Optional Components 14 Power Connections 17 Emergency Shutdown Procedure 17 Data Interface Connection Ports 18 Gas Inlet Outlet Connections 19 Warning L...

Page 3: ...Menu 46 Time Files Tab 47 Calibration Tab 50 Laser Adjust Tab 52 MIU tab 54 Analog Output Tab 55 Service Tab 57 Shutting Down the Analyzer 58 4 Maintenance 59 Daily Operation Checklist 59 Mirror Ring Down Time and Maintenance 59 Replace the Fuse 60 Appendix A About Gas Analyzers and Laser Absorption Spectroscopy 63 Conventional Laser Absorption Spectroscopy 63 ABB LGR s Off Axis Integrated Cavity ...

Page 4: ... corrections Cybersecurity This product is designed to be connected to and to communicate information and data via a network interface It is your sole responsibility to provide and continuously ensure a secure connection between the product and your network or any other network as the case may be You shall establish and maintain any appropriate measures such as but not limited to the installation ...

Page 5: ...fications Symbols Standards Tested and Met 2004 108 EU EMC EN61326 1 Title 21 Code of Federal Regulations chapter 1 sub chapter J WEEE Directive The analyzer is not subject to WEEE Directive 2002 96 EC Waste Electrical and Electronic Equipment or relevant national laws e g ElektroG in Germany The product must be disposed of at a specialized recycling facility Do not use municipal garbage collectio...

Page 6: ...r runs safely without risk to the operator Modifying the analyzer to operate with the case cover open can injure personnel Bypassing the analyzer interlock switch to open the case cover during analyzer operations can cause serious bodily injury Even though the analyzer provides a second layer of protection such as a laser cover to prevent the user from the invisible laser beam or any secondary ref...

Page 7: ...aintenance instructions and maintenance kits Make sure the analyzer is unplugged before working with the internal components Failure to do so may result in damage to the analyzer and electric shock Safety Provisions for a Chemical Spill Follow these precautions when dealing with all chemicals Keep all chemical containers away from heat sparks and open flames Use only on grounded equipment and with...

Page 8: ...u type in fields and also button choices that you enter Warning Icons Table 2 shows and describes the warning icons used in this manual Table 2 Warning Icon Descriptions Icon Meaning NOTE or IMPORTANT Emphasizes facts and conditions important to analyzer operation WARNING General Warning Icon gives general safety information that must be followed to avoid hazardous conditions WARNING Electrical Wa...

Page 9: ...emperatures between 20o C and 60o C Save the original shipping materials to use when returning the analyzer to ABB LGR if factory service or repair is needed Table 3 lists and describes the safety icons on ABB LGR shipping boxes Follow these instructions when transporting and storing boxed analyzers Table 3 Transportation and Storage Icon Descriptions Icon Meaning Store your analyzer in a sheltere...

Page 10: ...ng transportation to ABB LGR is at customer s expense 5 No charge repair parts may be sent at ABB LGR s sole discretion to the purchaser for installation by purchaser 6 ABB LGR s liability is limited to repair or replace any part of the analyzer free of charge if ABB LGR s examination discloses the part to be defective The laws of some locations may not allow the exclusion or limitation on implied...

Page 11: ...sued RMA number must be prominently displayed on the return package No returns will be accepted collect or C O D On all warranty returns ABB LGR will pay the shipping charges on the return of the merchandise to the customer Customer Support ABB provides product support services worldwide To receive product support either in or out of warranty contact the ABB office that serves your geographical ar...

Page 12: ...Ambient Humidity 99 relative humidity non condensing Operating Temperature 5 45o C Maximum Altitude 6 000 Feet Power Requirements 60 Watts typical 24 30 VDC battery power 66 Watts typical 15 230 VAC 50 60hz AC adaptor 120 Watts maximum Fuse Ratings 32 VAC 10 Amps Fast blow fuse Cable Plugs and Voltage for EC Countries See page 105 Always use the power supply cord provided by ABB LGR See page 105 f...

Page 13: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 13 External Dimensions 18 5 W x 14 D x 7 H Weight 15 kg Cell Volume 345 cc cavity 10 cc tubing Total volume 355 cc ...

Page 14: ...ower cable o 25 pin control signal cable Maintenance kit MTN KIT 64 This recommended kit includes o Rebuild kit for internal pump o Mirror cleaning kit o 32GB solid state hard drive DC power pack with battery ACC DCCASE B DC power case no battery ACC DCCASE Battery hook up kit ACC UP 12V Ultra portable analyzer back pack ACC UP BP Wireless user interface with iPad ACC WIFI iPad Wireless user inter...

Page 15: ...stem failure The hard drive fails or is corrupted Battery power to the analyzer is interrupted If any of these failures occur you will need a mouse keyboard and monitor to restart the analyzer A back up hard drive is recommended If you have not received all of these components contact ABB LGR at icos support ca abb com ...

Page 16: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 16 Figure 2 shows the left side of the analyzer Figure 2 Left Panel Figure 3 shows the right side of the analyzer Figure 3 Right Panel ...

Page 17: ...onnector and the power switch Figure 4 Power Connections and AC Voltage Selection Switch The analyzer is equipped with a three prong power plug The third prong of the facility electrical outlets must be grounded Failure to ground the third prong may result in electrical damage to the analyzer and electrical shock to the operator Emergency Shutdown Procedure If the analyzer malfunctions and require...

Page 18: ...nnects the analyzer to a local area network LAN and allows access to the data directory using an external computer Serial port 9 pin D sub For real time digital measurement output Video port 15 pin D sub Connects an external monitor to the analyzer Analog out port Provides a DC voltage proportional to the measured gas concentration If this output is connected to an external device it must be termi...

Page 19: ...ET port An external In Line filter is provided and can be attached to the INLET port During normal operation the pump draws the sample through the analyzer inlet and exhausts the sample through the waste port The acceptable inlet gas pressure range is 0 to 10 psig The push connect WASTE port may be Routed to the facility ventilation system or Remain open expelling non hazardous gas into the room a...

Page 20: ...ves a description of the warning labels Figure 7 shows the location of the labels on the analyzer Table 4 Warning Labels and Descriptions Label Description The laser is rated Class 3 invisible laser radiation when the housing is open Only trained maintenance personnel may open the analyzer housing General laser warning label Figure 7 Warning Label Locations ...

Page 21: ...gure 5 for a detailed description of the connections Connect the Inlet Outlet Plumbing Connections 1 If applicable connect a sample tube not provided from the push connect INLET port on the right panel of the analyzer to your sample source 2 The provided External In Line filter can be used for additional filtering by connecting to the push connect INLET port on the right panel of the analyzer Figu...

Page 22: ...d Battery power to the analyzer is interrupted If any of these failures occur you will need a keyboard mouse and monitor to restart the analyzer A back up hard drive is recommended 2 Click on the launch button to manually launch the analyzer a The launch button is the abbreviated name of the gas analyzer For example the GLA151 N2OM1 button displays N2O CH4 Figure 11 b If you don t make a selection...

Page 23: ...e the integrity check runs The integrity check runs for one to two minutes before launching the analyzer s control software Do not turn off the computer while the integrity check is running Figure 10 File System Integrity Check Screen Thermal Stabilization Run the analyzer for four hours before collecting data This allows the internal temperature to stabilize The exact final cell temperature will ...

Page 24: ... auto launch countdown to manually start recording measurements by clicking the launch button o The launch button is the abbreviated name of the gas analyzer For example the GLA151 N2OM1 button displays N2O CH4 Figure 11 Open the auto launch window by clicking Service Turn off the analyzer by clicking Shutdown Figure 11 shows the Launch Service screen Figure 11 Launch Service Screen ...

Page 25: ...he Service button on the Launch Service screen From this interface you can Change the auto launch delay timing Transfer files from the internal hard drive to an external storage device connected via USB by clicking Files Restore the analyzer s factory settings by clicking Restore Figure 12 shows the Auto Launch screen Figure 12 Auto Launch Screen ...

Page 26: ...n on the User Interface Control Bar Figure 13 Figure 13 Control Bar Security Button 2 For initial login use the default Linux credentials for the username and password Figure 14 as follows User lgr Password 3456789 If you change and forget this password you will not be able to recover it without a factory restore There is only one Linux account Figure 14 Login Dialog Box 3 Click Login ...

Page 27: ...lyzer is displayed at the bottom of the main panel Green The analyzer is functioning properly Yellow The data may not be reliable or maintenance is required soon Red The analyzer requires maintenance to correct an identified fault Refer to the Alarm Status Display section on pages 31 33 for detailed Temperature Status and Analyzer Status descriptions This panel contains the User Interface Control ...

Page 28: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 28 User Interface Control Bar Use the control bar to operate the analyzer Figure 16 User Interface Control Bar ...

Page 29: ...ts the rate at which data is written to the log file Figure 22 Parameter Window Displays the Time Current time Data File Current filename to log data Gas Temperature Temperature in Cell Celsius C Gas Pressure Pressure in Cell Torr Laser A τ Laser A ring down time micro seconds µs Laser B τ Laser B ring down time micro seconds µs o Only applicable for two laser systems MIU Multiport Inlet Unit Rate...

Page 30: ...isplay The four displays within the display function are Numeric Alarm Status Spectrum TimeChart This section describes the displays Numeric Display This is the default display It appears when the analyzer is first turned on or re initialized Figure 17 shows an example of the Numeric Display screen for the GLA151 N2OM1 It displays the numeric readout of the last measurements of gas at a specific c...

Page 31: ...l status of the analyzer The Alarm Status is color coded Green The analyzer is functioning properly Yellow The data may not reliable or maintenance is required soon Red The analyzer requires maintenance to correct an identified fault Figure 18 shows the Alarm Status Display with all parameters functioning properly Figure 18 Alarm Status ...

Page 32: ... operating temperature range 11 CROSS OVER 14 Analyzer Temp Temperature High Low Warning The temperature is the high warning set point The temperature is the low warning set point 17 Fault NaN reading Occurs when there is a false or undefined value NaN not a number 19 Dead Band 20 Acceptable Range No warning alarm No warning alarm If the Alarm Status is Yellow or Red please refer to the Maintenanc...

Page 33: ...s when linelock control voltage is no longer able to control 9 Signal Power A B Signal power has degraded Occurs when laser signal power has degraded by 20 of the factory value 10 Maintenance Maintenance needed now Occurs when the analyzer requires maintenance every 381 days 11 CROSS OVER 12 Data Health A B Fit is not optimal The laser fitting condition is not optimal Occurs when residuals of fit ...

Page 34: ... UM ICOS EN REV B 2 34 19 Dead Band 20 Performance No warning alarm No warning alarm If the Alarm Status is Yellow or Red please refer to the Maintenance section on page 59 If issue continues please contact icos support ca abb com ...

Page 35: ...ot shows the corresponding optical absorption displayed as black circles and the peak fit resulting from signal analysis as a blue line Figure 19 and Figure 20 show the Spectrum Displays for the following analyzers GLA151 N2OM1 GLA151 N2OCM GLA151 N2OM1 Portable Methane Nitrous Oxide Analyzer The measured N2O concentration is shown in parts per billion ppb and CH4 and H2O in parts per million ppm ...

Page 36: ...GLA151 N2OCM Portable Nitrous Oxide Analyzer Carbon Monoxide Analyzer The measured N2O and CO concentrations are shown in parts per billion ppb and H2O in parts per million ppm on the bottom of the Spectrum Display Figure 20 Spectrum Display GLA151 N2OCM ...

Page 37: ... Chart with a continuous flow of gas A 10 point running average in black is shown going through the raw data shown in blue Click on the drop down box in the lower right corner of either window to toggle between available gases and to adjust the number of significant figures The data is saved to the file indicated in the left corner of the parameter window Figure 21 TimeChart Display ...

Page 38: ...the User Interface Control Bar Figure 16 is selected a pop up box appears to allow rate control adjustments to the operating mode and plot frequency Figure 22 Figure 22 Rate Control Screen The Operating Mode radio buttons allow you to change the rate at which data is written to the log file To adjust the rate 1 Click the Rate button clock icon on the User Interface Control Bar Figure 16 The Data R...

Page 39: ... 22 2 Click the Plot on Demand radio button to manually plot the data a When selected the Refresh Plot button appears on the Main Panel display When Refresh Plot button is selected Figure 21 current data is added to the Main Panel display 3 Click the Plot every Nth fit radio button to automatically set the rate at which the data is updated on the Main Panel display a For example if you set the val...

Page 40: ...ext ASCII format and contain labeled columns displaying The time stamp of each recorded measurement Gas concentration Cell pressure Torr Cell temperature Celsius Ambient Temperature Celsius Ring Down Time microseconds The format can be changed in the Time Files menu of the Setup panel Figure 30 Figure 23 shows a typical data file for the GLA151 Series QC portable analyzer Figure 23 The Beginning o...

Page 41: ...re 24 Figure 24 File Transfer Menu Local Hard Drive left pane and USB Flash Drive right pane 4 Transfer data files from the analyzer hard drive to a USB storage device by dragging and dropping the files from the hard drive pane to the USB device pane Use the left mouse button to highlight one or multiple files in the window The directory windows default to the local hard drive on the left screen a...

Page 42: ...ring files 5 Click the Unmount USB button Wait for the Safe to Remove USB Memory Device message before removing the USB memory device 6 Click Close to exit the File Transfer Menu Removing the USB memory device before seeing the Safe to Remove pop up message may result in loss of data ...

Page 43: ... Figure 24 creates a daily folder containing new data files for each day that the analyzer operates To access the data files for a specific date double click the folder Each file from that day is displayed in chronological order Figure 25 Each file is a single zipped txt file using the following convention XXX_YYYY MM DD_f0000 txt zip Examples of files in the daily directory are shown in Figure 25...

Page 44: ...rganized by date Figure 26 To access the archived files double click the Archive folder Figure 24 Each file is a single zipped txt file using the following convention YYYY MM DD zip Each zipped file contains the data files for the day that the analyzer operated Examples of files in the archive directory are shown in Figure 26 Figure 26 Archive Directory ...

Page 45: ...EV B 2 45 File Transfer Error Screen The File Transfer Error screen Figure 27 displays when The USB Key does not have enough storage space The device is not recognized Try again with a correctly inserted USB device Figure 27 File Transfer Error ...

Page 46: ...additional configurations and services To enter Setup mode 1 Click the Setup button on the User Interface Control Bar Figure 28 Figure 28 Setup button on the User Interface Control Bar 2 The default Time Files screen is displayed Figure 29 Figure 29 Setup Menu Tabs with Time Files Screen Selected ...

Page 47: ...ate the analyzer to a local gas standard Enable the laser offset adjustment Configure the optional Multi Port Inlet Unit Configure the Analog Output Service screen for technicians to check on the status of the analyzer Use these function tabs to make adjustments to the analyzer and its operation Time Files Tab The Time Files menu allows you to adjust the time zone manually set the clock adjust the...

Page 48: ...e data files The available timestamp formats are shown in Table 7 New file creation intervals when running continuously can be set by adjusting the value in the Output Interval minutes spinner control box Table 7 Available Time Stamp Formats Time Stamp Name Format Absolute Local American mm dd yyyy hh mm ss sss Absolute Local European dd mm yyyy hh mm ss sss Absolute GMT American mm dd yyyy hh mm ...

Page 49: ...arity Stop Bits The actual rate of the serial output is equal to the Logged File Rate i e 1 Hz divided by the Rate specified in the Time Files menu Use a null modem serial cable to connect the analyzer serial port to an external computer About The About section displays analyzer specific information such as the Build Date of the current software Version of the code IP Address Serial Number of the ...

Page 50: ...elect the Calibration tab at the top of the screen to enter the Calibration menu Figure 31 Figure 31 Calibration Setup Screen 3 On the top right panel of the screen under Reference Gas Settings select the checkbox next to the gas you wish to calibrate 4 Enter the known concentration for your local gas standard 5 Connect your reference gas supply to the push connect inlet port on the left side of t...

Page 51: ...ibration Flow 9 When the Calibration Complete message is displayed click the CLOSE button 10 Enter TimeChart by selecting the Display button on the User Interface Control Bar and verify that the displayed concentration correctly corresponds to your local gas standard 11 Repeat steps 1 10 for each gas you wish to calibrate The time of latest calibration is stored in Reference Gas Settings within th...

Page 52: ...ive drift Figure 33 Laser adjustment may be needed for the following reasons The laser s wavelength has drifted beyond the target range of the analyzer The analyzer is operated outside the recommended temperature range Figure 33 shows the offset between absorption peaks and target lines Figure 33 Absorption peaks off of target lines Laser voltage adjustment needed ...

Page 53: ...age using the arrow buttons to shift the peaks until they are centered on their respective target lines a Up Arrow Peaks adjust to the right b Down Arrow Peaks adjust to the left 5 Deselect the Disable Laser Frequency Lock check box The software resumes automatic tracking and control of the laser wavelength 6 Click Close to exit the menu and return to the Main Panel Figure 34 shows the laser volta...

Page 54: ...red configuration These ports are directed to the inlet port of the analyzer for sampling unknown gases and reference gases The MIU menu can be configured to control which gases are introduced to the analyzer and for how long Figure 35 See the MIU Appendix on page 86 for detailed instructions on configuring and controlling the MIU Figure 35 Control Menu for the Optional Multi Port Inlet Unit MIU 8...

Page 55: ...r example Set 5 Volts 10 ppm on the expectation that the gases measured will be in the range near 2 ppm with occasional bursts up to almost 10 ppm Set 5 Volts 5 ppm to get exactly two times greater sensitivity on the analog output with the expectation that the concentration will not go above 5 ppm If the measured concentration goes above the maximum expected value for the Analog Output the on scre...

Page 56: ...nments D Sub Pins Pin Assignment Output Voltage 1 N2O Signal 0 5 Volts 2 CH4 Signal 0 5 Volts 3 H2O Signal 0 5 Volts 4 Empty 5 Empty 6 N2O Ground 0 7 CH4 Ground 0 8 H2O Ground 0 9 Empty Table 9 describes the pin assignments for the GLA151 N2OCM Table 9 D Sub Pins and Pin Assignments D Sub Pins Pin Assignment Output Voltage 1 N2O Signal 0 5 Volts 2 CO Signal 0 5 Volts 3 H2O Signal 0 5 Volts 4 Empty...

Page 57: ...ndent and are set based upon the last fixed setting Figure 38 Service Screen The Service tab contains 3 Service buttons Reset Tau Power button resets the mirror ringdown time and laser power to current settings This is typically done after mirrors have been cleaned Reset Counter button resets the of days that maintenance is required This is typically done after yearly maintenance Launch TeamViewer...

Page 58: ...n Figure 40 Figure 39 User Interface Control Bar Exit Button 3 Click the OK button to halt data acquisition close the current data file and display the shutdown screen Figure 40 Figure 40 Analyzer Shutdown Prompt 4 When the You may turn off the instrument message displays Figure 41 you can safely shut off power to the analyzer by pushing the OFF switch on the left panel of the analyzer Figure 2 Fi...

Page 59: ...ow system An increase in pressure could be indicative of a leak in the system or a pump failure Every 3 6 days Check the Laser Offset and adjust if necessary Figure 33 Mirror Ring Down Time and Maintenance Measurement cell mirrors are protected from contamination by an internal inlet filter With continued use the mirrors may gradually decline in reflectivity If a significant change occurs in the m...

Page 60: ...place the fuse 1 Flip the Power switch to OFF on the left side of the analyzer Figure 4 2 Unplug the 11 30 VDC power connector from the left side of the analyzer Figure 4 3 Unlatch the Case Securing Latches Figure 42 a Push the latch button b Pull down on the latch Figure 42 Case Securing Latches 4 Open the case top Figure 43 Figure 43 Open the Case Top ...

Page 61: ...ICOS EN REV B 2 61 5 The fuse is located on the upper portion of the analyzer on the left side above the internal pump Figure 44 Figure 44 Analyzer Fused Inlet 6 Pull the fuse out of the fuse fixture Figure 45 Figure 45 Remove the Fuse ...

Page 62: ...gure 46 Figure 46 Install the New Fuse in the Fuse Holder 8 Carefully close the case top and push the latches to lock the case 9 Re attach the 11 30 VDC power connector into the left side of the analyzer Figure 4 10 Flip the Power switch to the ON position on the left side of the analyzer Figure 4 ...

Page 63: ...ere Iv is the transmitted intensity through the sample at frequency Io is the reference laser intensity prior to entering the cell S is the absorption line strength of the probed transition L is the optical path length of the laser beam through the sample 𝛘 is the mole fraction P is the gas pressure Φv is the line shape function of the transition at frequency 𝞾 In this case If the laser line width...

Page 64: ...ngth Line strength of the probed transition Enables you to determine the mole fraction directly from the relation Use this equation to determine gas concentrations even in hostile environments without using calibration gases or reference standards These values are measured Mixtures containing several species Flows at elevated temperatures and pressures ...

Page 65: ...ctive optical path length is 2500 meters Figure 48 Schematic Diagram of an Off Axis ICOS Analyzer Because the path length depends only on optical losses in the cavity and not on a unique beam trajectory like conventional multi pass cells or cavity ring down systems the optical alignment is very robust allowing for reliable operation in the field The effective optical path length is determined rout...

Page 66: ...Protocol request when the analyzer is initialized If the analyzer does not receive a reply the analyzer Disables the Ethernet port Does not attempt another DHCP request until the analyzer is restarted When both conditions are met the data directory can be accessed using a Windows computer on the same LAN To access the data directory 1 Click Start Run and enter the IP address of the analyzer Exampl...

Page 67: ...al Area Network LAN is not available plug the analyzer into a standalone broadband router example Netgear Model RP614 to enable the analyzer to obtain a Dynamic Host Configuration Protocol DHCP address from the router when the analyzer is started Then plug any Windows computer into the same broadband router to access the data directory A crossover Ethernet cable will NOT allow an external computer...

Page 68: ...www tp link com support and enter the TP Link model number TL WR802N You must restart the analyzer whenever you change router modes for the mode change to take effect Local Connection You can also bypass the wireless router if you want to connect the analyzer to a local network Refer to the Connect Analyzer to Local Network section on page 72 for instructions on this configuration Wireless Control...

Page 69: ...ptop connect to the router using the SSID and password on the router Figure 49 Figure 49 Router SSID and Password Location 2 On the same device launch a Web browser then type http tplinkwifi net in the address bar Figure 50 Figure 50 Logging In 3 Enter admin in lowercase for both the username and password 4 Click Login ...

Page 70: ... ICOS INSTRUCTIONS UM ICOS EN REV B 2 70 5 On the left panel click Quick Setup Figure 51 Figure 51 Start Router Configuration 6 Click Next 7 In the Main Panel select the Access Point button Figure 51 8 Click Next ...

Page 71: ... or AP Wireless Password click Next To change the default Wireless Network Name and or AP Wireless Password change the names in the Wireless Network Name and or AP Wireless Password fields then click Next Figure 52 Wireless Setting Screen 10 In the Network Setting screen click Next Figure 53 Figure 53 Network Setting Screen 11 Click Finish 12 Restart the analyzer ...

Page 72: ...ernet Union 2 Remove the Ethernet union next to the router Figure 54 3 Plug the black cable into either port of the Ethernet union Figure 55 Figure 55 Plug Cables into Ethernet Union 4 Plug the blue cable into the other port of the Ethernet union Figure 55 5 Connect a local area network LAN cable from an external computer to the analyzer Ethernet port located on the external side panel ...

Page 73: ... software on the device you are connecting to the analyzer Set up VNC Software on Android Devices 1 On the Android device go to Settings WiFi Connect to Wireless Network 2 Connect to the wireless SSID network listed on the router sticker Enter the TP Link wireless router example TP LINK 775C as shown in Figure 49 For ultraportable analyzers the TP Link wireless router is installed inside the analy...

Page 74: ...a Wireless devices can compete for dynamic addresses If the 192 168 100 100 address does not connect then use 192 168 100 101 8 Record the IP address of the Android device because it will be necessary to refer to it in Step 12 9 Install the VNC software by searching and installing from the Google Play store Search for Android vnc viewer and install the application by tapping on the Install button ...

Page 75: ...igure 60 Figure 60 VNC Application Installation Setup Screen 12 In the Address field enter the address of the analyzer 192 168 100 100 or 192 168 100 101 that you recorded in Step 8 The IP address of the analyzer will be whichever address the Android device is not For example if the IP address of the Android device that was displayed in Step 8 is 192 168 100 101 then the IP address of the analyzer...

Page 76: ...button to connect the Android device to the analyzer The analyzer software interface screen displays on the device The screen size is adjustable to fit the screen of the device Figure 61 Figure 61 Analyzer Software Interface Display with Size Adjustment for Android Devices ...

Page 77: ...the analyzer and may be accessed by opening the case For all other analyzers the optional TP Link wireless router is attached to the outside of the case 3 Select your SSID network For example TP LINK D036 Figure 62 Figure 62 Network Connections Screen 4 The Enter Password screen appears Figure 63 In the Password field enter the wireless password on the router sticker Figure 49 5 Select Join Figure...

Page 78: ...ecord the IP address of the iOS device because it will be necessary to refer to it in Step 12 Figure 65 Device IP Address Confirmation Screen 9 Install the VNC software by searching and installing it from the App store Mocha VNC Lite for iOS is the software used in this example Figure 66 An Internet connection is required for this step Complete instructions for installing Mocha VNC Lite for iOS ca...

Page 79: ...e 68 Figure 68 Mocha VNC Lite Configure Screen 12 Enter the analyzer s address in the VNC server address field 192 168 100 100 or 192 168 100 101 from Step 8 The IP address of the analyzer will be whichever address the iOS device is not For example if the IP address of the iOS device that was displayed in Step 8 is 192 168 100 101 then the IP address of the analyzer will be 192 168 100 100 13 In t...

Page 80: ...ess Figure 69 Figure 69 Setup Configurations Screen 15 To connect the iOS device to the analyzer tap the IP Config you set up The analyzer software will display on the device Figure 70 The screen size is adjustable to fit the screen of the device Figure 70 Analyzer Software Interface Screen Size Adjustment for iOS Devices ...

Page 81: ...ws device open Wireless Router options 2 Locate the sticker on the router Figure 49 3 Click on the Wireless Network Connections icon in the bottom left of the screen Figure 71 to open the Windows Wireless Networks dialog box Figure 72 Figure 71 Wireless Connections Icon Figure 72 Windows Wireless Networks ...

Page 82: ...ple TP LINK 775C to display the Connect to a Network dialog box Figure 73 5 In the Security key field enter the wireless password located on the router sticker Figure 49 6 Click OK Figure 73 Network Connections Security Screen 7 The Connection Status dialog box displays Figure 74 Figure 74 Current Connectivity Screen ...

Page 83: ...he device is connected through the wireless router by selecting the router Figure 75 Figure 75 Wireless Network Connection Screen 9 Verify the IP address of the Windows device a Right click on the TP LINK 775C network connection b Click Status Figure 76 Figure 76 Current Connectivity Screen ...

Page 84: ...10 The Wireless Network Connection Status dialog box displays Figure 77 Figure 77 Wireless Network Connection Status Window 11 Click the Details button to display the Network Connection Details window Figure 78 Figure 78 Network Connection Details Window ...

Page 85: ... www realvnc com products vnc documentation 5 0 guides user Cha pter1 html 14 Open the program by clicking the Connect button Figure 79 Figure 79 Real VNC Viewer Installation Screen 15 Enter the analyzer s address in the VNC server address field 192 168 100 100 or 192 168 100 101 from Figure 78 The IP address of the analyzer will be whichever address the Windows device is not For example if the IP...

Page 86: ...er and for how long By sampling references periodically during an ongoing data run you can post correct the data for long term drift when active calibration cannot be done ABB LGR offers two versions of the MIU 8 port MIU 8 16 port MIU 16 Figure 80 shows the front panel of a 16 port MIU Figure 80 16 Port MIU Front Panel Control of the MIU is unidirectional The analyzer does not receive feedback on...

Page 87: ... the Components 1 Connect the provided power cable into the fused power entry module on the back panel 2 Connect the 25 pin control cable from the MIU to the TO MIU port on the back panel of the analyzer 3 Connect a 1 4 Teflon tube from your gas source into one of the numbered inlet ports Repeat for multiple gases 4 When connecting the tubing push the tube into the port until you feel a click in o...

Page 88: ...menu to specify what ports are sampled and for how long Figure 82 shows the Gas Manifold Control Screen for the MIU not yet enabled Figure 82 Gas Manifold Control Screen for the MIU not yet enabled 3 Populate the unknown gas valve sequence a Valve The current valve being sampled corresponds to the port number on the MIU b Seconds How long the analyzer should sample the gas in seconds c Description...

Page 89: ...to the port number on the MIU b Seconds How long the analyzer should reference the gas in seconds c Description Input a short text description associated with the reference gas connected to that valve 5 Click on the Start with reference gas valve sequence check box if you wish to run your reference gases first 6 Use the arrow scroll bar to select the number of times to run the unknown gas sequence...

Page 90: ... current valve being sampled referenced and its text description is shown in the parameter window of the User Interface Control Bar Figure 84 The description is Displayed on the parameter window of the User Interface Control Bar during analysis Figure 84 Saved to a data file Figure 84 User Interface Control Bar showing MIU information 9 When sampling is complete disable the MIU by returning to the...

Page 91: ...ts are included as part of the recommended maintenance kits MTN CLEAN M See Optional Components If further maintenance is required contact ABB LGR for service o Technical Support icos support ca abb com Only authorized persons may open the analyzer cover or perform internal maintenance Contact ABB LGR for maintenance instructions and mirror cleaning kits Make sure the analyzer is unplugged before ...

Page 92: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 92 Figure 85 shows the mirror cleaning kit Figure 85 Mirror Cleaning Kit ...

Page 93: ...ticularly carpet fibers are attracted to freshly cleaned mirrors and can re contaminate the mirror before it can be replaced onto the cavity 2 Power off the analyzer and remove the power plug 3 Wait 10 minutes to allow the pressure in the measurement cell to reach ambient room pressure 4 Unlatch the Case Securing Latches Figure 86 a Push the latch button b Pull down on the latch Figure 86 Unlatch ...

Page 94: ...e case top Locate the cavity as shown in Figure 87 Figure 87 Ultra Portable Case Open 6 Remove the safety shield by un screwing the two flat head screws and carefully lifting the shield off of the laser assembly Figure 88 Figure 88 Remove the Laser Safety Shield ...

Page 95: ... 89 Figure 89 Remove the screws from the Detector Bracket 8 Slide the detector bracket off of the mirror flange Figure 90 shows the analyzer with the detector removed and the mirror flange exposed 9 Without removing any wires from the detector place the detector assembly as far from the cavity as possible Figure 90 Open Ultra Portable Analyzer Case Showing Mirror Flanges ...

Page 96: ...plug the wires by separating the connectors as shown in Figure 91 The detector flange does not have this connector Figure 91 Separate Connectors 11 Remove both mirror flanges by unscrewing and loosening the three large flange bolts The bolts are captive and cannot be removed from the flange Figure 92 Figure 92 Remove the Mirror Flanges ...

Page 97: ...3 Place the flange on a clean dust free surface as shown in Figure 93 Figure 93 Mirror with Reflective Side Down 14 Detach the backing ring by loosening and removing the three small screws using the hex driver from the Chapman screwdriver kit Figure 94 Figure 94 Detach the Backing Ring If the small diameter face seal O ring sticks to the mirror gently remove it and place the O ring back into the g...

Page 98: ...TRUCTIONS UM ICOS EN REV B 2 98 15 Place the mirror flange on top of the Teflon tube provided and carefully push the flange downwards until the mirror pops out of the flange O ring Figure 95 Figure 95 Releasing the Mirror ...

Page 99: ...nts to the high reflectivity HR side of the mirror The opposite side of the mirror is coated with an anti reflection coating and is referred to as the AR side Figure 96 Mirror in Cleaning Block with High Reflectivity Side Down 2 Remove any loose particles of dust using the compressed gas canister by gently blowing across the mirror from a distance of six to eight inches Do not attempt to remove pa...

Page 100: ...he interior of the bottle and discard the rinse acetone 5 Fill the Acetone Dropper Bottle approximately full with fresh acetone Figure 97 6 Repeat steps 3 through 5 for Methanol Figure 97 Acetone and Methanol Dropper Bottles 7 Separate a single sheet of lens tissue from the pack Place the sheet on the mirror with the mirror at the edge of the sheet Leave enough tissue free on the near side of the ...

Page 101: ...e of the mirror Figure 99 Figure 99 Drag Wipe Step 1 9 Pull the tissue towards you with a consistent but light pressure over the full length of the mirror at a speed that matches the evaporation rate of the solvent from the surface This is called drag wiping Figure 100 Figure 100 Drag Wipe Step 2 If you see liquid on the mirror after you pass the wipe over it then you are wiping too fast ...

Page 102: ...e drag wipe process 13 Discard the tissue after each drag wipe 14 Invert the mirror in the mirror cleaning block so that the high reflectivity HR surface is on top and the arrow is facing up Figure 101 Figure 101 Clean the HR Surface 15 Perform the drag wipe process in this order on the high reflectivity side a Acetone 4 drops b Methanol 4 drops c Acetone 4 drops d Methanol 4 drops ...

Page 103: ...ing equal pressure on the mirror Repeat this until all screws are tight Be careful not to over tighten the screws as this may strip the threads 4 Place the flange onto the cavity and tighten the large bolts until they are tight Be careful not to over tighten the screws as this may strip the threads 5 Reconnect the wire cable on the front flange to the analyzer connector Figure 91 6 Place the Laser...

Page 104: ...rt 2003 0001 30ml LDPE Dropper Bottle Nalgene Part 2411 0030 Acetone HPLC Plus 99 9 Sigma Aldrich Part 650501 Methanol HPLC Plus 99 9 Sigma Aldrich Part 646377 Compressed air duster canister McMaster Carr Part 7437K35 Chapman screwdriver kit McMaster Carr Part 5399A16 Edmond Scientific lens tissue Part NT52 105 ...

Page 105: ... ICOS INSTRUCTIONS UM ICOS EN REV B 2 105 Appendix G Cables Table 11 describes the power cables shipped with your analyzer Table 11 Power Cables Region Cable Specifications Australia and New Zealand United Kingdom ...

Page 106: ...USER MANUAL ICOS INSTRUCTIONS UM ICOS EN REV B 2 106 Europe United States ...

Page 107: ...egard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without...

Reviews: