Initialize and Run the Analyzer
To initialize the analyzer:
1.
Press the power switch on the left side of the analyzer to the ON position. (0 = OFF / - =
ON)
The internal computer initializes, and a screen (Figure 9) displays as the program loads.
The
Launch Service screen displays after initialization. (Figure 11)
The wireless interface will not work if:
The analyzer has a system failure
The hard drive fails or is corrupted
Battery power to the analyzer is interrupted
If any of these failures occur, you will need a keyboard, mouse, and
monitor to restart the analyzer. A back-up hard drive is recommended.
2.
Click on the launch button to manually launch the analyzer.
a. The launch button is the abbreviated name of the gas analyzer. For example,
the GLA151-N2OM1 button displays N2O-CH4. (Figure 11)
b. If you don’t make a selection within 120-seconds, the analyzer automatically
defaults to the
Main Panel Numeric display. (Figure 17)
3. Click on the maintenance SERVICE button (Figure 11) if you need more time or need to
choose a maintenance setting. (Figure 12)
Figure 9: Start-up Screen in Busy Mode